Alarm conditions
Alarms put the pump in Attention mode.
A (ALARM)
This alarm shows the letter A followed by two numbers. A-alarms cause all insulin delivery to stop. Your
pump settings are retained. If this alarm repeats often, call our 24 Hour HelpLine for assistance.
AUTO OFF
Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature,
and so insulin delivery has been stopped.
BATT OUT LIMIT
Occurs if the battery has been out of the pump for more than five minutes. Verify that the pump time and
date are correct. If not correct, go to the UTILITIES MENU and reset the date and time.
BOLUS STOPPED
If this alarm occurs, it is very important to check bolus history to review how much of the bolus was
actually delivered. If necessary, reprogram a bolus with the amount not yet delivered.
BUTTON ERROR
Occurs if a button has been continually pressed for more than three minutes.
CHECK SETTINGS
When this alarm is active, you should check and/or reprogram your pump settings, including the time/
date.
Troubleshooting and alarms 137