User's Manual

Table Of Contents
DRAFT
Micropilot FMR53, FMR54
12 V. 1, Rev. 7, 26-10-2012 Endress+Hauser
Output
Output signal HART
Signal coding FSK ±0.5 mA over current signal
Data transmission rate 1 200 Bit/s
Galvanic isolation Yes
PROFIBUS PA (in preparation)
Signal coding Manchester Bus Powered (MBP)
Data transmission rate 31.25 kBit/s, voltage mode
Galvanic isolation Yes
FOUNDATION Fieldbus (in preparation)
Signal coding Manchester Bus Powered (MBP)
Data transmission rate 31.25 kBit/s, voltage mode
Galvanic isolation Yes
Switch output
For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output.
Switch output
Function Open collector switching output
Switching behavior Binary (conductive or non-conductive), switches when the programmable switch point is
reached
Failure mode non-conductive
Eectrical connection values U = 10.4 to 35 V
DC
, I = 0 to 40 mA
Internal resistance
R
I
< 880 W
The voltage drop at this internal resistance has to be taken into account on planning the
configuration. For example, the resulting voltage at a connected relay must be sufficient
to switch the relay.
Insulation voltage floating, Insulation voltage 1 350 V
DC
to power supply aund 500 V
AC
to ground
Switch point freely programmable, separately for switch-on and switch-off point
Switching delay freely programmable from 0 to 100 sec. , separately for switch-on and switch-off point
Number of switching cycles corresponds to the measuring cycle
Signal source
device variables
Level linearized
Distance
Terminal voltage
Electronic temperature
Relative echo amplitude
Diagnostic values, Advanced diagnostics
Number of switching cycles unlimited