Use Instructions

Instructions For Use | 5352 | © 2018 Endotronix Inc.
100474-00 Rev 1
Environment
The CalEQ may not meet its performance specifications if stored or used
outside the temperature and humidity ranges listed below.
Temperature
Operation: 5 – 35 °C (41 – 95 °F)
Storage: -20 – 60 °C (-4 – 140 °F)
Relative humidity
Operation: 25 – 85% (non-condensing)
Storage: 25 – 85% (non-condensing)
Classification
Class I Equipment
• IP Rating
o Computer Monitor: IP54 Front Panel
o Barcode Scanner: IP42
o Reader: IP22
o Docking Station: IP21
• Continuous Use
Safety Testing
Meets all relevant parts of IEC 60601-1 Ed. 3.1
Electrical Characteristics
Power
Multiple Socket Outlet (MSO)
o Input: 230V
, 50Hz, 13A (UK) or 16A (EU)
o Output: 230V , 50Hz, 13A (UK) or 16A (EU)
o Cord Length: approx. 3 meters
• Computer Monitor
o System Input: 12 – 48V
o AC Adapter
Input: 100 – 240V , 50/60 Hz, 1.3 – 0.6A
• Output: 19V 4.74A, 90W MAX
Cord Length: approx. 2 meters
'HÿQLWLRQRI6\PEROV
The following symbols are used on the labels of the Cordella Pulmonary
Artery Sensor System.
Manufacturer’s catalogue or part number so that the
medical device can be identified.
Manufacturer’s batch code so that the batch or lot
can be identified.
Manufacturer’s serial number so that a specific
medical device can be identified.
Authorized representative in the European
Community.
Need for the user to consult the instructions
for use.
Medical device that needs protection from light
sources or heat.
Temperature limits to which the medical device can
be safely exposed.
Humidity limits to which the device can be safely
exposed.
Range of atmospheric pressure to which the device
can be safely exposed.