User`s guide
Magnification Calibration
A diffractogram will be calculated (Figure 2-13).
Figure 2-13 Diffractogram
• Follow the on screen instructions (Figure 2-14)
A red line will appear on the diffractogram.
• Position the endpoints of the red line on two symmetrical diffraction
peaks (Figure 2-13).
• Press OK to specify the reciprocal unit and the d-spacing (in the
corresponding real units) in the next window (Figure 2-15).
Figure 2-14 High magnification calibration instructions
UltraScan User’s Guide Reference Series Rev 1.0 2-12