User`s guide

Magnification Calibration
Position it on a feature of known size.
Press OK on the “Calibrate image” window.
Enter the correct distance for the selected feature (for example 10 line
pairs of cross grating sample where the distance = 10 x 0.463µm) in the
“Calibration” window and select the units (see Figure 2-9).
Figure 2-9 Calibration window
Press OK.
Choose YES in the next window (see Figure 2-9) to complete the
magnification calibration.
Figure 2-10 Choose YES to complete the calibration
The calibration can be checked on the calibration table containing pairs of
value: the nominal microscope magnification and the calibrated value. To view
the magnification table:
Select “Magnification table” (Figure 2-11) from the “Microscope”
menu (Figure 2-6).
UltraScan User’s Guide Reference Series Rev 1.0 2-10