Camera Configuration
Figure 2-3 (a) Camera Menu (b) Camera configuration
• Select Configure Camera in the Camera menu (Figure (2-3a).
• In the Camera configuration dialog box (Figure 2-3b), you can change
the name of the camera by typing in the Name field.
• Image rotation can be specified in the Rotation pull-down menu.
• The coordinates for column defect correction are entered in the Defects
field. For example, entering the following:
r0-8, r4088-4095, c0-8,c4088-4095
(r and c indicates row and columns respectively) corrects for the 8
pixels wide edge which does not contain active CCD pixels in the case
of the UltraScan 4000.
• It is important to specify the Primary and Alternate (if there is one set
up) shutter available to the camera.
• Finally you should specify the shutter status, i.e. the state of the shutter
when the CCD camera is not acquiring any images. Shutter Normally
Closed is the default setting.