Entering Signal Limits (cont.)
2. Setting the Transducer Signal High Limit
Enter the high limit for the transducer signal coming into the
TransPort. This value is set to 85 by default.
• Press [ENT] to accept the current setting, or
• Use the numeric keys to enter a new value and press [ENT].
Acceptable values are between 25 to 100.
3. Setting the Correlation Peak
This prompt asks you to select the acceptable low limit for the
transducer signal quality. This value is set to 100 by default.
• Press [ENT] to accept the current value, or
• Use the numeric keys to enter a new value and press [ENT].
Acceptable values are between 0 to 500.
The E4: SIGNAL QUALITY error message appears if the signal
quality falls outside the set limit.
4. Entering the Sound Speed Limit
Enter the acceptable limits for the sound speed based on the
conditions in your system. This value is set to 20% by default.
• Press [ENT] to accept the current value, or
• Use the numeric keys to enter a new value and press [ENT].
Acceptable values are between 1 to 50 percent.
The E2: SOUND SPEED ERROR message appears if the
calculated fluid sound speed exceeds the fluid sound speed
entered in the SYSTeM menu by more than this limit.