User`s manual

Programming Site Data 3-31
December 2003
Setting Up Analog Outputs (cont.)
2. Enter the base (zero) value for the analog output. This value
represents the 4 mA/0 mA output (in flow units). Use the numeric
keys to enter a value.
3. Press [ENT].
4. Enter the full scale value for the analog output. This value
represents the 20 mA output (in flow units). Use the numeric keys
to enter a value.
5. Press [ENT].
6. If necessary, repeat this procedure for OUT2.
You have completed entering data into the Input/Output option.
You may at this time do one of the following:
Use the [] and [] arrow keys to verify or change data.
Continue to enter data in other menus as described in this
section.
Save the entered parameters as described on page 3-32.
Press the [EXIT] key to begin taking measurements. If you
press [EXIT] without saving, the TransPort gives you the
opportunity to save. Refer to page 3-32 to save the site data.
If you decide to exit, you must have already programmed the
system and pipe parameters in order to make flow measurements.