3-30 Programming Site Data
Setting Up Analog Outputs
The TransPort can send a current signal to your recording device
in different ranges. The first prompt for the analog outputs, OUT1
and OUT2, asks you to select the type of range you want to
output. Press [F4] to select OUT 1 or scroll to the next screen to
select OUT 2 ([F1]).
Use the appropriate feature key to select one of the following:
• OFF (Disable Analog Output)
• 0-20m (0-20 mA)
• 4-20m (4-20 mA)
Note: If you are not using the analog outputs, select [F1] to turn
them off.
The next series of prompts enables you to set up the analog output
by selecting the units and entering base and full scale values. Use
the following steps to set up the analog output:
1. Select the analog output type. Use the appropriate feature key
to select one of the following:
• VELocity
• VOLUMetric
• +TOTaL(Forward)
• -TOTaL (Reverse)
• TIME
• POWER (only with the energy option on)
• +ENeRGy (only with the energy option on)
• -ENeRGy (only with the energy option on)
• DIAGnostics
Note: Units for analog output are the same as those available in
the Run mode.