3-26 Programming Site Data
Entering Parameters for Wetted Transducers (cont.)
9W. Entering the Calibration Factor
The Calibration Factor is used as a reference for the flowmeter
electronics to ensure accurate readings. Generally, if you enable
the Reynolds Correction Factor, the calibration factor should be
set to 1.00. Otherwise the typical factor is between .90 and 1.00.
Use the numeric keys to enter a value and press [ENT].
You have completed entering data into the Pipe Menu. You may
at this time do one of the following:
• Use the [↑] and [↓ ] arrow keys to verify or change data.
• Continue to enter data using other Program Menu options as
described in this section.
• Save the entered parameters as described on page 3-32.
• Press the [EXIT] key to begin taking measurements. If you
press [EXIT] without saving, the TransPort gives you the
opportunity to save. Refer to page 3-32 to save the site data.
If you decide to exit, you must have already programmed the
system parameters in order to make flow measurements.