User`s manual
Programming Site Data 3-21
December 2003
Entering Parameters for Clamp-On Transducers
(cont.)
12C. Display of Transducer Spacing
The next prompt displays the suggested spacing of the
transducers calculated from the information entered. Record this
number and use it to properly space transducers. Use the Liquid
Transducer Installation Guide as a guide to installing the
transducers on the pipe.
Note: The spacing calculated by the meter can be overwritten
(using the numeric keys) if necessary to match the actual
physical spacing of the transducers. Do not change the
spacing by more than ±10% from the S calculated by the
meter.
You have completed entering data into the PIPE option. You may
at this time do one of the following:
• Use the [↑ ] and [↓] arrow keys to verify or change data.
• Continue to enter data using other Program Menu options as
described in this section.
• Save the entered parameters as described on page 3-32.
• Press the [EXIT] key to begin taking measurements. If you
press [EXIT] without saving, the TransPort gives you the
opportunity to save. Refer to page 3-32 to save the site data.
If you decide to exit, you must have already programmed the
system parameters in order to make flow measurements.