User`s manual
Screen and Error Messages G-5
December 2003
Log Active, END
only
hit any key
This message appears when in the LOG
Menu. This indicates the Log is still
compiling data. You may only edit the
End Time
Log DONE, to
inspect
hit any key
This message appears when in the LOG
Menu. This indicates the Log is com-
plete. Hit any key and the TransPort
displays the log on the screen.
Low Signal This message appears while using the
thickness gage. This indicates the input
device signal is poor.
NVR FAIL This message appears at power up. This
message indicates that the non-volatile
RAM has failed. Contact the factory.
NVR O.K. This message appears at power up. This
message indicates that the non-volatile
RAM has passed.
Outside limits,
value rejected.
This messages appears when in the
CAL Menu. This message indicates that
the calibration for the analog output is
invalid. Hit any key to clear this mes-
sage. The TransPort defaults to the last
valid calibration for the analog output.
This message also appears if you do not
have an ammeter or a multimeter con-
nected to the TransPort while trying to
calibrate the analog output
__ range is X.XX
to X.XX
This message appears while in the
PROGRAM Menu. This indicates the
entered number is unacceptable. Hit
any key and enter a number within
range.
S This message appears in the indicator
box. S indicates the shift key function is
activated. Press a key or the shift key
twice to deactivate the shift function.
Message Description/Action