User`s manual
Screen and Error Messages G-3
December 2003
E1: Low Signal This message appears on the message
line on the screen. This indicates that
the ultrasonic signal is poor and
exceeds the limits programmed in the
PROGRAM Menu SETUP option.
Check the transducer as described on
page 10-12.
E2: Sound Speed
Error
This message appears on the message
line on the screen. This indicates that
the sound speed for your fluid exceeds
the limits programmed in the PRO-
GRAM Menu SETUP option. Refer to
Sound Speeds and Pipe Size Data to
look up the sound speed. Also, refer to
page 10-3.
E3: Velocity Range This message appears on the message
line on the screen. The velocity exceeds
the programmed limits (velocity limits
are programmed using the PROGRAM
Menu SETUP option). Refer to page
10-3.
E4: Signal Quality This message appears on the message
line on the screen. This indicates the
signal quality exceeds the limits pro-
grammed in the PROGRAM Menu
SETUP option. Check transducer and
couplant as described on page 10-12.
Relocate transducers of possible. Also,
refer to page 10-3 for more information.
E5: Amplitude
Error
This message appears on the message
line on the screen. This indicates the
amplitude exceeds the limits pro-
grammed in the PROGRAM Menu
SETUP option. Refer to page 10-4.
Message Description/Action