User`s manual
Entering Special Transducer Pipe Parameters A-15
December 2003
Entering Parameters for Wetted Transducers (cont.)
8W. Enter the Axial Length (L)
GE Panametrics has calculated the axial length (L) of the
ultrasonic signal based on the transducer configuration for your
particular application. Find the axial dimension on the flowcell or
from other supplied documentation.
1. Press the appropriate feature key to select the desired units.
2. Use the numeric keys to enter the axial length. Press [ENT].
Note: If the axial length is not supplied, refer to Appendix B to
measure the axial length.
9W. Selecting Tracking Windows
If you want tracking windows:
• Press [F1] for no or
• Press [F2] for yes.
10W. Entering the Fluid Type
Enter the fluid type. You may choose one of the following:
• water
• oil
• methanol
• ethanol
• LN
2
• freon
• other