User`s manual

Entering Special Transducer Pipe Parameters A-11
December 2003
Entering Parameters for Clamp-On Transducers
(cont.)
16C. Display of Transducer Spacing
The next prompt shows the suggested spacing of the transducers
calculated from the information entered. Use the Liquid
Transducer Installation Guide as a guide to installing the
transducers on the pipe with the spacing dimension (S) given
here.
Set the transducers to the spacing shown at this prompt.
Note: The spacing calculated by the unit can be overwritten
(using the numeric keys) if necessary to match the actual
physical spacing of the transducers.
You have completed entering data into the PIPE option. You may
at this time do one of the following:
Use the [] and [] arrow keys to verify or change data.
Continue to enter data in other menus as described in Chapter
3, Programming Site Data.
Press [EXIT] to exit the Program Menu and begin taking
measurements.
If you decide to exit, you must have already programmed the
system parameters in order to make flow measurements.