User`s manual
Using the Ultrasonic Thickness Gage 9-11
December 2003
Programming the Ultrasonic Thickness
Gage (cont.)
Figure 9-6: Thickness Gage Program Screen
Low Signal Threshold
This parameter sets the signal strength alarm. If the signal
strength falls below this value, a “Low Signal” message appears
on the screen. This number should be set to 45 unless otherwise
instructed by the factory. To change this number:
1. Use the numeric keys to enter the new number.
2. Press [ENT].
The TransPort proceeds to the next prompt.
Detection Threshold
This number represents the percent of peak the TransPort uses to
make measurements. The TransPort will consider anything above
the entered percentage as part of the signal. Consult the factory
before changing this number. If you are instructed to change this
number, use the steps below:
1. Use the numeric keys to enter the new number.
2. Press [ENT].
The TransPort proceeds to the next prompt.
PROGRAM Start
LOW SIGNAL THRESHOLD
45