9-10 Using the Ultrasonic Thickness Gage
Programming the Ultrasonic Thickness
Gage
Once you access the Thickness Gage option, the TransPort
converts from an ultrasonic flowmeter to an ultrasonic thickness
gage. The thickness gage has a Program Menu that enables you to
view or change the following parameters:
Low Signal Threshold
Detection Threshold
Transducer Delay
Gain Level
Signal Inversion
The thickness gage settings are entered at the factory. You should
not change them unless instructed by the factory. The only setting
you may change without affecting the thickness gage operation is
the system units.
To enter the Thickness Gage Program Menu:
1. Press the [CAL] key to enter the Calibration Menu.
2. Press [F3] to select “WALL.”
3. Press the [PROG] key.
The screen appears similar to the one shown in Figure 9-6 below.
The indicator box shows you are at the beginning of the Program
Menu for the ultrasonic thickness gage.
If you only want to change the System Units, press the [ENT] key
until the “System Units” prompt appears, then follow the
instruction in the corresponding section.