9-8 Using the Ultrasonic Thickness Gage
Calibrating the Thickness Gage
The ultrasonic thickness gage is factory calibrated; therefore, you
only need to perform calibration if you receive new transducers
or have a reason to doubt the accuracy of the readings.
To calibrate you will need:
a test block, which is shipped with the thickness transducer (or
test block of known sound speed and length)
ultrasonic couplant
To calibrate the thickness gage, you must enter the Calibration
Menu and then enter the appropriate data.
Entering the Calibration Menu
1. Press the [CAL] key to enter the Calibration Menu.
2. Press [F3] to select “WALL.”
3. Press the [CAL] key.
Enter Calibration Data
1. Use the numeric keys to enter either one of the test block
lengths, then press [ENT]. If you are using a GE Panametrics
supplied test block, the length is printed on the block.
2. Use the numeric keys to enter the other test block length, then
press [ENT]. If you are using a GE Panametrics supplied test
block, the length is printed on the block.
Note: To perform a one point calibration, enter the same
number for both lengths.