User`s manual

Calibrating Analog Outputs and Inputs 8-7
December 2003
Low and High Calibration (cont.)
You have completed calibrating the analog inputs. You may now
do one of the following:
If you are finished calibrating inputs/outputs, press [EXIT]
twice and the TransPort begins taking measurements.
If you want to calibrate the analog output, press [EXIT] and
refer to page 8-2 for instructions.
To reconnect the analog outputs and inputs, refer to Chapter 2,
Initial Setup.