TransPort™ PT868 Portable Flowmeter
December 2003 Process Control Instruments TransPort™ PT868 Portable Liquid Flowmeter User’s Manual 910-122C
December 2003 Warranty Each instrument manufactured by GE Panametrics is warranted to be free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE Panametrics. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser.
December 2003 Return Policy If a GE Panametrics instrument malfunctions within the warranty period, the following procedure must be completed: 1. Notify GE Panametrics, giving full details of the problem, and provide the model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE Panametrics will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided. 2.
December 2003 Table of Contents Chapter 1: Features and Capabilities Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1 System Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5 Chapter 2: Initial Setup Making Electrical Connections. . . . . . . . . . . . . . . . . . . . . .2-1 Charging and/or Replacing Batteries . . . . . . . . . . . . . . . . .
December 2003 Table of Contents (cont.) Chapter 5: Printing Data Setting Up the Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2 Printing “Live” Measurements . . . . . . . . . . . . . . . . . . . . . .5-3 Printing Logged Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5 Printing Site Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6 To Stop Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6 Examples of Printed Data . . . .
December 2003 Table of Contents (cont.) Chapter 11: Specifications Overall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-2 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-3 Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-6 Pipe Size and Material . . . . . . . . . . . . . . . . . . . . . . . . . . .11-7 Available Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
December 2003 Chapter 1 Features and Capabilities The TransPort™ Model PT868 is a transit-time flowmeter which combines all the features of a full-size flowmeter with the advantages of a portable instrument. This section describes the TransPort features and general system, and explains the theory of operation. Overview The TransPort measures acoustically conductive single-phase fluids, which may have a second phase present.
December 2003 The TransPort has the ability to store site data for up to 20 sites into files which can be accessed at a later time. A menu-driven user program asks you all the necessary setup questions for a particular site. Once you complete answering the necessary questions, you simply save the information into a file. The TransPort stores these files and other data in non-volatile memory, which retains the information even if power is off.
December 2003 System Description The TransPort is one part of the flowmeter system. The flowmeter system consists of two essential subsystems: the flowcell and the electronics package (the TransPort). The Flowcell The flowcell is that part of the system that uses ultrasonic pulses to interrogate the flow. It consists of the flowcell pipe and the transducers. A.
December 2003 Electronics Package The TransPort consists of circuits that generate, receive, and measure the travel time of the ultrasonic pulses. It also contains a microcomputer that controls operation and calculates flow measurement parameters. Specific circuits function as follows: A.TRANSMIT SIGNAL GENERATOR - The transmit signal generator, under control of the microcomputer and timing circuit, synthesizes the signal that drives the transmitter. B.
December 2003 Theory of Operation The TransPort is a transit-time ultrasonic flowmeter. When ultrasonic pulses are transmitted through a moving liquid, the pulses that travel in the same direction as the fluid flow (downstream), travel slightly faster than the pulses that travel against the flow (upstream). The TransPort uses various digital signal processing techniques, including cross-correlation, to determine transit times and then uses the transit times to calculate flow velocity.
December 2003 Chapter 2 Initial Setup Before making measurements, you must prepare the TransPort for operation.
December 2003 PWR XDCR UP- DOWN- STREAM STREAM 1 ANALOG I/O RS232 1 7 2 3 2 6 FLOW 1 PWR RS232 1. 12VDC-28VDC 6W MAX. 2. RETURN 1. GND(BROWN) 2. TX(RED) 3. RX(ORANGE) 4. DTR(YELLOW) 5. No Connection . (GREEN) 4 6 5 8 ANALOG I/O 1. 4-20mA/0-10V ANALOG OUT 2. 16V (switched) 3. SUPPLY TEMP 4. RETURN TEMP 5. GND 6. AOUT2 7. GND 8. No Connection 6.
December 2003 Analog Input/Output Connections The TransPort provides one 4-20 mA or 0-20 mA current output and two 4 to 20-mA analog inputs with switchable 16-V supply for loop powered temperature transmitters. Connect the inputs/ outputs using a LEMO® multi-pin connector as shown in Figure 2-1. The pin numbers for the connector and the color code for the standard input/output cable are shown in Table 2-1.
December 2003 Charging and/or Replacing Batteries The TransPort comes with self-contained, built-in rechargeable batteries to support portable operation. For optimum performance, these batteries require a minimum of maintenance. Charging the Batteries When you receive the TransPort, you will need to initially charge the batteries. When fully charged, the batteries provide 6 to 8 hours of continuous operation.
December 2003 Replacing the Batteries If for any reason you need to replace the rechargeable batteries, replace with 4-C size NiCad high energy rechargeable batteries (GE Panametrics Part Number 200-044). To replace the batteries, open the panel located on the back of the TransPort unit, disconnect the batteries, and replace with new ones (see Figure 2-2).
December 2003 Replacing the Batteries (cont.) Caution! Replace batteries only with the specified rechargeable NiCd batteries. The battery charges when the unit is off. Do not attempt to recharge non-rechargeable batteries.
December 2003 Powering On and Off To operate the TransPort, the power cord must be plugged into line voltage or the battery must be charged as described previously. To turn the TransPort on, press [ON] which is in the upper-righthand corner of the keypad. Immediately upon power up the TransPort displays the GE Panametrics logo and the software version.
December 2003 Powering On and Off (cont.) After the meter conducts all the self checks, the screen then appears similar to the one shown in Figure 2-3. To turn the TransPort off, press [SHIFT] (red key) and [ON]. SITE 1 * VELOCITY Ft/s 12.
December 2003 Using the Screen and Keypad The essential features for operating the TransPort are the screen and keypad. Although these features are common, there are characteristics about each feature that need to be explained. Screen The primary function of the screen is to display information in order for you to accurately and easily make measurements.
December 2003 Screen (cont.) The middle of the screen is the work area, which displays measurements, graphs, menu prompts, etc. In addition, this area also displays error code messages, which are described in more detail in Chapter 10, Diagnostics and Troubleshooting. Indicator Box SITE 1 * VELOCITY Ft/s 12.50 Work Area VEL VOLUM +TOTL Feature Key Options -TOTL Figure 2-4: The TransPort Screen The bottom of the screen also displays the four feature key options.
December 2003 Keypad The TransPort keypad has 30 keys and each key has a second function. The functions for each key are as follows (see Figure 2-5 on the next page): • 4 feature keys ([F1], [F2], [F3], [F4]) - enable you to select the special functions or parameters which appear at the bottom of the screen. • 12 numeric keys (including - and .) - enable you to enter numeric data. • Arrow Keys>4 arrow keys ([← ], [→], [↑ ], [↓ ]) - enable you to move through the menu options.
December 2003 Keypad (cont.) To execute a shift function, press [SHIFT] (notice an “S” appears in the indicator box to show the shift key is activated), then the desired key. When a key is pressed after shift is activated, the shift function is canceled. When you press the shift key twice, it locks and an “SL” appears in the indicator box. To unlock the [SHIFT] key, press it once more (the “SL” symbol disappears).
December 2003 Obtaining On-Line Help The TransPort offers on-line help screens that describe various features. You can access on-line help at any time by pressing [SHIFT] and [HELP]. When you enter help, the screen displays on-line help pertaining to the task you are performing. For example, if you enter help while setting up the data logger, help for the data logger appears. Use the four feature key options to view on-line help details.
December 2003 Using the Backlight To provide more visibility for the display, the TransPort can be equipped with a backlight. If your unit has a backlight, you can operate it as described below. To turn on the backlight: Press and hold down the [ON] key, then press the [7] key. To brighten the back light, press the [7] key repeatedly. To turn the backlight off: Press and hold down the [ON] key, then press the [4] key.
December 2003 Chapter 3 Programming Site Data The TransPort has a Program Menu that enables you to enter information that is specific to each site, such as transducer types, installation configurations, pipe material, etc. The TransPort requires that you enter only the system and pipe data in order to make measurements (shown in the Quick Start Flow Measurement Guide).
December 2003 Entering the Program Menu To enter the Program Menu, press the [PROG] key on the keypad. The screen should appear similar to that shown in Figure 3-1. Refer to Appendix D, Menu Maps, Figure D-1 for a flow diagram of the Program Menu.
December 2003 Entering the Program Menu (cont.) • COMMunication - enables you to choose the baud rate and stop bits for a monitoring device. • SETUP - enables you to set signal limits, response time, and decimal digits for each site. Use the following sections to program the desired information into the TransPort. Remember, after you enter the desired information, refer to page 3-32 to save the site data.
December 2003 Entering the ACTIV Mode Note: It is not necessary to enter the ACTIV menu to operate the PT868, as the flowmeter automatically runs in TransitTime mode. The first option in the Program Menu is ACTIV, which enables you to select the Transit-Time mode of measurement. (On a PT868-R flowmeter, the option enables you to choose between the Transit-Time and Transflection modes. The Transflection option is not available for the PT868.
December 2003 Entering System Parameters The second option in the Program Menu is SYSTeM, that enables you to select and enter global information for TransPort operation. You must enter these parameters in order to make flow measurements. (These parameters were also discussed briefly in the Quick Start Flow Measurement Guide.) To enter the SYSTeM option, press [F2]. The screen appears similar to Figure 3-3. The indicator box indicates you are at the beginning of the SYSTeM option.
December 2003 Entering System Parameters (cont.) Use the steps that follow to enter the system parameters. 1. Choosing the Energy Option The first prompt asks you if you want to measure energy. Use the appropriate feature key to respond: • Press [F1] if you do not want to measure energy flow rate, or • Press [F2] if you do want to measure energy flow rate. If you choose to measure energy flow rate, the TransPort will ask additional questions within this menu.
December 2003 Entering System Parameters (cont.) 4. Turning the Stopwatch Totalizer On or Off The Stopwatch Totalizer function enables you to either totalize all liquid flow continuously, or to manually start and stop totalizing liquid flow. • Press [F1] to select automatic totalizing, or • Press [F2] to select manual control. If you select manual control, a”T” appears in the upper right corner of the screen. To stop the totalizer in manual mode during operation, press [ON] and the digit 1.
December 2003 Table 3-1: Volumetric Unit Options (Continued) English Volumetric Units Metric Volumetric Units ft3/m - Cubic Feet per Minute m3/m- Cubic Meters per Minute ft3/h- Cubic Feet per Hour Mm3/h - Millions of Cubic Meters per Hour Mft3/d- Millions of Cubic Feet per Day Mm3/d - Millions of Cubic Meters per Day BBL/S - Barrels per Second BBL/S - Barrels per Second BBL/M - Barrels per Minute BBL/M - Barrels per Minute BBL/H - Barrels per Hour BBL/H - Barrels per Hour MBBL/D - Millions of
December 2003 Entering System Parameters (cont.) 6. Choosing Totalizer Units To select the units for totalized flow: a. Use the [←] [→] keys to view all the options available. See Table 3-2 for a list of available totalizer units. b. Press the appropriate feature key to make a selection. Table 3-2: Totalizer Units Options English Totalizer Units Metric Totalizer Units Gal - U.S. Gallons L - Liters MGAL - Mega U.S.
December 2003 Entering System Parameters (cont.) 8. Choosing Total Energy Units Use the appropriate feature key to select the units for total energy. • Press [F1] for kBTU - Thousands of British Thermal Units • Press [F2] for MMBTU - Millions of British Thermal Units • Press [F3] for kWHr - Kilowatts-Hours • Press [F4] for Tons - Tons. 9. Choosing Heating or Cooling System Use the appropriate key to select whether you are making measurements in a heating or cooling system.
December 2003 Entering System Parameters (cont.) 12.Setting the Time The last prompt in SYSTeM asks you to set the time. You can accept the currently displayed time by pressing [F1], or change it as described below: a. Press [F2] to edit the time. b. Use the numeric keys to enter the hour. c. Press the appropriate feature key to select AM or PM, and press [ENT]. d. Use the numeric keys to enter the minutes, and press [ENT]. e. Use the numeric keys to enter the seconds, and press [ENT].
December 2003 Entering Pipe Parameters The third option in the Program Menu is PIPE, that enables you to enter information for the pipe characteristics and the transducers. You must enter these parameters in order to operate the TransPort. These parameters were also discussed briefly in the Quick Start Flow Measurement Guide supplied separately. The series of prompts that appear depend on the type of transducers you are using.
December 2003 Entering Pipe Parameters (cont.) PIPE PROGRAM PROGRAM Pipe parameters Start TRANSDUCER NUMBER 40 TRANS Figure 3-4: Screen After Pressing PIPE Key 1.
December 2003 Entering Parameters for Clamp-On Transducers If you are using clamp-on transducers, the following prompts will appear: • • • • • • • • • • • Wedge Temperature Input (for medium or high temperature) Pipe Material Pipe O.D. (pipe outside diameter) Pipe Wall Thickness Lining (material and thickness) Tracking Windows Fluid Type Reynolds Correction Factor (and Kinematic Viscosity) Calibration Factor # of Traverses Transducer Spacing 2C.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 3C.Entering the Pipe Material Since clamp-on transducers send their ultrasonic signals through the pipe wall, the sound speed of the pipe material must be factored in to obtain an accurate flow measurement. Clamp-on transducers are applicable on a variety of pipe materials, only a few of which are among the options for this prompt.
December 2003 Entering Parameters for Clamp-On Transducers (cont.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 4C.Entering the Outside Diameter of the Pipe If you do not know the outside diameter (OD) of your pipe, use either the on-line help or consult Sound Speeds and Pipe Size Data that contains some standard pipe sizes with their corresponding outside diameters. For greater accuracy, measure the pipe OD at the location where the transducers will be mounted. The PT868 can also calculate the outside diameter from the pipe circumference.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 6C.Entering Pipe Lining The next prompt asks you to enter the pipe lining material. Clamp-on transducers can be used on pipes with a variety of pipe linings; however, only a few are among the available options. If your lining is not among the options, select “OTHER” and enter the sound speed of your lining material from the tables in Sound Speeds and Pipe Size Data. a.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 8C.Entering the Fluid Type The next prompt asks you to enter the fluid type.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 9C. Entering the Reynolds Correction Factor The Reynolds Correction Factor is a number based on the Kinematic Viscosity and flow rate of the fluid. Generally, it should only be enabled when using transducers mounted in a diametrical path (such as clamp-on or tilted diameter).
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 12C. Display of Transducer Spacing The next prompt displays the suggested spacing of the transducers calculated from the information entered. Record this number and use it to properly space transducers. Use the Liquid Transducer Installation Guide as a guide to installing the transducers on the pipe.
December 2003 Entering Parameters for Wetted Transducers If you are using wetted transducers, you will get the following prompts: • • • • • • • • Pipe OD (pipe outside diameter) Pipe Wall Path Length (P) Axial Dimension (L) Tracking Windows Fluid Type Reynolds Correction Factor (and Kinematic Viscosity) Calibration Factor 2W. Entering the Outside Diameter of the Pipe The first prompt for wetted transducers is pipe outside diameter (OD).
December 2003 Entering Parameters for Wetted Transducers (cont.) 3W. Entering the Pipe Wall Thickness Enter the thickness of the pipe wall. If you are not sure of the pipe wall thickness, use either the on-line help or consult the brochure Sound Speeds and Pipe Size Data that contains some standard pipe sizes with their corresponding wall thicknesses. For higher accuracy, the pipe wall thickness can be determined by using the optional internal ultrasonic thickness gage as described in Chapter 9.
December 2003 Entering Parameters for Wetted Transducers (cont.) 5W.Enter the Axial Length (L) Enter the axial length (L) of the ultrasonic signal. GE Panametrics has calculated the axial dimension based on the transducer configuration for your particular application. Find the axial dimension on the flowcell or from other supplied documentation. a. Press the appropriate feature key to select the desired units. b. Use the numeric keys to enter the axial length. Press [ENT].
December 2003 Entering Parameters for Wetted Transducers (cont.) Note: If you have selected the energy option, only three fluid types will appear on the menu: water, mixed (water, glycol mixture) and other. If you select “OTHER,” the TransPort prompts you to enter the sound speed of the fluid. If you have enabled the tracking windows, you may choose the following: • Water, 0-100° C • Water, 0-260° C • Oil • Other — Enter sound speed (c3) limits Use the appropriate feature key to make a choice.
December 2003 Entering Parameters for Wetted Transducers (cont.) 9W. Entering the Calibration Factor The Calibration Factor is used as a reference for the flowmeter electronics to ensure accurate readings. Generally, if you enable the Reynolds Correction Factor, the calibration factor should be set to 1.00. Otherwise the typical factor is between .90 and 1.00. Use the numeric keys to enter a value and press [ENT]. You have completed entering data into the Pipe Menu.
December 2003 Entering Input/Output Parameters — I/O The fourth option in the Program Menu is Input/Output. This menu enables you to enter information to set up the analog inputs and outputs. In addition, this menu enables you to select how the TransPort will handle the outputs during error and low flow conditions. To enter the Input/Output option, press [F4]. The screen appears similar to Figure 3-5. Refer to Appendix D, Menu Maps, Figures D-1E, D1-F, D1-G and D1-H, for flow diagrams of the I/O option.
December 2003 Selecting Error Handling The first setting, ERROR, asks you to select how the flowmeter will handle the analog outputs in the event of a fault condition. Use the appropriate feature key to select one of the following: • • • HOLD - which holds the last good reading LOW - which forces the reading low (0 mA or 4 mA) HIGH - which forces the reading high (20 mA) The next prompt asks you to set the zero cutoff.
December 2003 To Set Up for Active Temperature Inputs: 1. Press [F2] to select IN 1, the supply input. 2. Select [F1] for ACTVE. 3. Enter the temperature for the supply input (base) that corresponds to a 4 mA signal, then press [ENT]. 4. Enter the temperature for the supply input (full) that corresponds to a 20 mA signal, then press [ENT]. 5. Press [F2] to select IN2, the active temperature return input. 6. Select [F1] for ACTVE. 7.
December 2003 Setting Up Analog Outputs The TransPort can send a current signal to your recording device in different ranges. The first prompt for the analog outputs, OUT1 and OUT2, asks you to select the type of range you want to output. Press [F4] to select OUT 1 or scroll to the next screen to select OUT 2 ([F1]).
December 2003 Setting Up Analog Outputs (cont.) 2. Enter the base (zero) value for the analog output. This value represents the 4 mA/0 mA output (in flow units). Use the numeric keys to enter a value. 3. Press [ENT]. 4. Enter the full scale value for the analog output. This value represents the 20 mA output (in flow units). Use the numeric keys to enter a value. 5. Press [ENT]. 6. If necessary, repeat this procedure for OUT2. You have completed entering data into the Input/Output option.
December 2003 Saving Site Data The fifth option in the Program Menu is SAVE, which enables you to save the entered site data in non-volatile memory. The TransPort can store up to 20 sets of site data in files. File names can be up to five characters. To enter the SAVE option, use the [←] [→] keys to scroll until “SAVE” appears on the screen and press [F1]. The screen appears similar to Figure 3-6. Refer to Appendix D, Menu Maps, Figure D-1I for a flow diagram of the SAVE option.
December 2003 Saving Site Data (cont.) To save site data: 1. Use the alphanumeric keys to enter a file name (up to five characters) or use a feature key to select a file to overwrite. 2. If you entered a new file name, press [ENT]. Your file is now stored in memory and remains resident until you retrieve another file. To check if the file is stored, use the [↑] key to scroll back to the “NAME” prompt, notice your file is assigned a feature key. You may have to use the [←] [→] to see your file.
December 2003 Recalling Files The sixth option in the Program Menu is RECaLL, which enables you to retrieve files that you saved earlier. To enter the RECaLL option, use the [←] [→] keys to scroll until “RECLL” appears on the screen and press [F2]. The screen appears similar to Figure 3-7. Refer to Appendix D, Menu Maps, Figure D-1J for a flow diagram of the RECaLL option. RECALL PROGRAM RECALL NAME SITE1 SITE1 SITE2 SITE3 SITE4 Figure 3-7: Screen After Pressing RECLL Key To recall a file: 1.
December 2003 Setting Up RS232 Communications Port The TransPort can transmit stored data and displayed readings using the RS232 communications port. The TransPort can also receive and execute commands remotely using a serial interface device such as a personal computer. To use the communications port, make connections as described in Chapter 2, Initial Setup (refer to Appendix C for more information on cables).
December 2003 Setting Up RS232 Communications Port (cont.) To set up communications parameters: 1. Choose the baud rate. Use the [←] [→] keys to view all the options. Then press the appropriate feature key to select the desired baud rate. 2. Enter the UART bits. Use the [←] [→] keys to view all the options. 3. Press the appropriate feature key to select the desired data and parity. The communications settings have now been selected. To exit the Program Menu, press [EXIT].
December 2003 Windows 3.X Systems (cont.) 4. Using the Transfer Menu, select the Receive Text File option. 5. A window appears asking for a file name. Select the desired directory and file name (with .prt extension) and then click on OK. 6. From the keypad on the PT868, enter the PRINT menu and select the log file to be uploaded to the computer in numeric format. See Chapter 4, Printing Data, for detailed instructions on this procedure. The logged data will begin appearing on the computer screen.
December 2003 Windows 9X/NT Systems (cont.) 5. After making any necessary changes, click on OK. 6. Open the TRANSFER menu and select CAPTURE TEXT. Enter the desired DRIVE:\DIRECTORY\FILENAME for the file about to be transmitted, and click on OK. 7. From the keypad on the PT868, enter the PRINT menu and select the log file to be uploaded to the computer in numeric format. See Chapter 4, Printing Data, for detailed instructions on this procedure. 8.
December 2003 Entering Setup Parameters The SETUP option contains four sub-options that enable you to set signal limits, response time, and decimal digits for each site, or to initialize the setup parameters to their default values. To enter the SETUP option, use the [←] [→] keys to scroll until “SETUP” appears on the screen, then press [F3]. The screen appears similar to Figure 3-9. Refer to Appendix D, Menu Maps, Figure D-1L for a flow diagram of the SETUP option.
December 2003 Entering Signal Limits The SIGNaL sub-option enables you to set the limits for the incoming signal and other parameters that affect the transducer signal. For example, you can adjust the low and high limits of the transducer signal strength. When the signal falls outside the programmed limits, the TransPort alarm sounds. To enter the SIGNaL sub-option, press [F1]. The indicator box shows you are at the beginning of the SIGNaL sub-option.
December 2003 Entering Signal Limits (cont.) 2. Setting the Transducer Signal High Limit Enter the high limit for the transducer signal coming into the TransPort. This value is set to 85 by default. • Press [ENT] to accept the current setting, or • Use the numeric keys to enter a new value and press [ENT]. Acceptable values are between 25 to 100. 3. Setting the Correlation Peak This prompt asks you to select the acceptable low limit for the transducer signal quality.
December 2003 Entering Signal Limits (cont.) 5. Entering the Low Limit for Velocity Enter the acceptable low limit for the velocity. • Press [ENT] to accept the current value (the default is -40.0 ft/s), or • Use the numeric keys to enter a new value and press [ENT]. Acceptable values are from -100 to 100 ft/s. 6. Entering the High Limit for Velocity Enter the acceptable high limit for the velocity. • To accept the current value, press [ENT]. The default is 40.0 ft/s.
December 2003 Entering Signal Limits (cont.) 8. Setting the Low Limit for the Amplitude Discriminator The next prompt asks you to enter the acceptable low limit for the amplitude discriminator. The amplitude discriminator measures the size of the transducer signal sent from the TransPort. This value is set to 14 by default. Acceptable values are between 0 and 100. • Press [ENT] to accept the current value, or, • Use the numeric keys to enter a new value and press [ENT].
December 2003 Entering Signal Limits (cont.) 10. Setting the Delta-T Offset Consult GE Panametrics before performing this step. Delta-T is the difference between the upstream and downstream transit time of the transducers. You usually need to enter a DeltaT offset when you are using a test flowcell provided by GE Panametrics. Acceptable values are between -1000 and 1000 nsec. To enter an offset, use the numeric keys to enter a value and press [ENT]. 11.
December 2003 Entering Signal Limits (cont.) 13. Setting the Transmitter Sample Size Consult GE Panametrics before performing this step. The transmitter sample size is the number of pulses each transducer (upstream and downstream) emits. This number is set to 8 by default. Use the appropriate feature key to respond. 14. Entering the Number of Errors Allowed The TransPort needs to know the number of errors it can record before it displays an error message.
December 2003 Entering Averaging Parameters The AVeRaGe sub-option enables you to select the response time. The TransPort enables you to select a response time for the unit to respond to a step change in flow rate. In general, the faster the response time, the less steady the reading will appear. Normally, use statistics for the best results. Statistics increases the averaging under steady flow conditions, but allows for a rapid response to step changes in flow rate.
December 2003 Chapter 4 Displaying Data The TransPort enables you to display data in various formats in order for you to compare and manipulate information. The TransPort has the following six options for the display format: • Big - shows one measurement at a time (see Figure 4-1). • Dual - shows two measurements simultaneously (see Figure 42). • Graph - shows one measurement at a time on an X/Y type line chart (see Figures 4-3 and 4-4).
December 2003 Big Format When you first power up, the TransPort automatically displays data using the Big Format as shown in Figure 4-1 below. SITE 1 VELOCITY Ft/s 12.50 VEL VOLUM +TOTL -TOTL Figure 4-1: Big Format Use the following steps to return to the Big Format from another display format: 1. Press the [DISP] key. 2. Press [F1] to select “BIG”. The TransPort automatically exits the Display Menu and displays the velocity. To display other measurements: 1.
December 2003 Dual Format The Dual Format enables you to show two measurements simultaneously as shown in Figure 4-2 below. SITE1 VELOCITY 12.5 VOLUMETRIC 33.7 VEL VOLUM +TOTL Ft/s GAL/M -TOTL Figure 4-2: Dual Format To select the Dual Format, press the [DISP] key, and press [F2] to select Dual. The TransPort automatically exits the Display Menu and displays two measurements. If you want to change the measurement displayed do the following: 1.
December 2003 Graph Format The Graph Format enables you to show velocity or volumetric flow on an X and Y line graph at specified time intervals as shown in Figure 4-3. The Graph Format is useful for indicating trends and changes in flow measurements. Use the sections below to set up and use the Graph Format. 15.2 GAL/M 09:00 V O L U M ST:10 NOV 92 09:34 START END Figure 4-3: Graph Format Setting Up the Graph Format Use the following steps to set up the Graph Format: 1.
December 2003 Graph Format (cont.) 4. Select the time increment for the X axis. Use the [←] [→] keys to view the following options: • • • • • • • 5 seconds 10 seconds 30 seconds 1 minute 3 minutes 6 minutes 12 minutes Press the appropriate feature key to make a selection. 5. At the next prompt, use the numeric keys to enter a value for the Y (vertical) axis maximum for the graph and press [ENT]. Note: Enter a Y axis value that is larger than the maximum expected flow. 6.
December 2003 Using the Graph Format Figure 4-4 below shows an example of a bidirectional flow graph.The Y axis represents the measurement value and the X axis represents the time. Note: The X axis for Positive only flow graphs is the bottom line of the graph. Status Line 15.2 GAL/M 09:38 V O Y Axis Cursor L X Axis U M ST:10 NOV 92 09:34 START END Message Line Figure 4-4: Graph Format The TransPort displays 120 points of data at a time.
December 2003 Using the Graph Format (cont.) Note: Since the time displayed on the status line will only show whole minutes and not seconds, if you selected a time increment of less than 1 minute, the time will not change every time you press the cursor key. For example, if you chose a 5-second increment, you will have to press the cursor key 12 times before the time changes. Use the [←] [→] keys to access the other feature key options for viewing the data in the graph format.
December 2003 Log Format The Log Format enables you to display logged data graphically or numerically. The TransPort displays all of the data in the log file; however, because of the limited space on the screen, the TransPort cannot display all data on the screen simultaneously. Therefore, you must use the feature key options to view the remaining data. Note: Before viewing a log, you must create a log. This process is described in Chapter 6, Logging Data.
December 2003 Using the Numeric Log Format Figure 4-5 below shows an example of a log displayed in the numeric format. The screen displays the measurement values numerically, including the time and date the measurement was taken. Each set of measurement values is called a record. The TransPort is capable of storing up to 120 pages of records. Each page consists of 120 records. The last line of the screen displays any error messages. Record Number Page Number PAGE# 1 REC #1 02:39:47 AM 1 JUN 94 -0.
December 2003 Displaying Logs in the Graphical Plot Format 1. Press the [DISP] key. 2. Press [F4] to select “LOG.” 3. Use the [←] [→] keys to view all the log file names. Use the appropriate feature key to select the desired file. 4. Press [F2] to select “PLOT.” 5. Enter a value for the Y (vertical) axis maximum (full scale) for the graph, then press [ENT]. Note: Enter a Y axis value that is larger than the maximum expected full scale. 6.
December 2003 Using the Graph Log Format Figure 4-6 shows an example of a bidirectional log graph. The Y axis represents the measurement value; the X axis represents time. Note: The X axis for Positive Only flow graphs is the bottom line of the graph. Status Line 12.5 Ft/s 09:00 V E L Y Axis Cursor X Axis ST:10 NOV 92 09:34 Message Line Figure 4-6: Log Graph Format The TransPort displays 120 data points on a page.
December 2003 Using the Graph Log Format (cont.) Note: Since the time displayed on the status line will only show whole minutes and not seconds, if you selected a time increment less than 1 minute, the time will not change every time you press the cursor key. For example, if you chose a 5-second increment, you will have to press the cursor key 12 times before the time changes. Use the [←] [→] keys to access the other viewing options. Each option is described below.
December 2003 Displaying the Signal The TransPort has a feature in the Display Menu that enables you to graphically display the transducer signal. This feature aids in troubleshooting transducer and other problems. To display the transducer signal, do the following: 1. Press the [DISP] key. 2. Press the [←] [→] keys until “SIGNL” appears, then press [F1]. The screen appears similar to that shown in Figure 4-7.
December 2003 Displaying the Signal (cont.) To view the signal, use the [] feature keys to move the cursor left and right within the current page. Use the [] feature keys to move quickly from page to page of the signal. When you move the cursor, the Status Line (see Figure 4-7 on the previous page) displays the signal amplitude and the time. Use the [←] [→] keys to access the other viewing options. Each option is described below.
December 2003 Displaying the Signal (cont.) When displaying the signal, the TransPort can display the following : • Upstream signal (Sup) • Downstream signal (Sdown) • Upstream Correlation (Cup) • Downstream Correlation (Cdown) • Cross-correlation (CC) You can toggle between the above displays using the [1] and [2] keys. [1] toggles up, and [2] toggles down. Additionally, [4] vertically enlarges the signal, and [5] reverts it to the original size.
December 2003 Chapter 6 Logging Data A powerful and flexible feature of the TransPort is data logging. The TransPort enables you to choose from up to six sets of parameters to log. You can also select the start time and date, end time and date, and time interval. Logs can run one at a time or simultaneously. Error and circular logs are also available. Logged data is internally stored in battery backed up memory. The TransPort stores logged data in pages, allocating 120 pages of memory for data logging.
December 2003 Once you make your data logging selections, the TransPort enables you to view the amount of memory left. You may also stop logging or view logged data while or after data is logged as described in Chapter 4, Displaying Data. This section describes: • How to enter the Log Menu. • How to setup the data logger. • How to view the data logger memory. • How to stop logging.
December 2003 Entering the Log Menu To access the Log Menu, press the [LOG] key. The screen should appear similar to that shown in Figure 6-1 below. The indicator box shows you are at the beginning of the Log Menu. Refer to Appendix D, Menu Maps, Figure D-3 for a flow diagram of the LOG Menu.
December 2003 Entering Data Log Parameters The first option in the Log Menu is STD. This option enables you to select the log parameters such as the measurements to log, how frequently to log them, and when to begin and end logging. To enter STD, press [F1]. The screen should appear similar to Figure 6-2 below. The indicator box shows that you are at the beginning of this option. NAME, MESSAGE, etc.
December 2003 Entering Data Log Parameters (cont.) 1. Entering a Log Name At this prompt, you may enter a new log name or choose to overwrite an existing log. • To set up a new log entry, use the alphanumeric keys to enter a file name (up to five characters) and press [ENT]. • To view or alter an existing log, use the [←] [→] keys to view all the log names and press the appropriate feature key to make a selection. 2. Entering a Log Message Enter a brief description (up to 21 characters) of the log.
December 2003 Entering Data Log Parameters (cont.) 5. Entering the Log Start Time Enter the time you want to begin logging. • Press [F1] to accept the displayed time and proceed to Step 6. • Press [F3] to start the log immediately and proceed to Step 7. • To enter a new time : a. Press [F2] to edit the start time. b. Press the appropriate feature key to select AM or PM. c. Use the numeric keys to enter the hour. Press [ENT]. d. Use the numeric key to enter the minutes. Press [ENT]. e.
December 2003 Entering Data Log Parameters (cont.) 7. Entering the Log End Time Enter the time you want to end the log. • Press [F1] to accept the current end time, and proceed to Step 8. Note: The log end time must exceed the start time by at least 5 minutes.If the log is circular, it will prompt you for a duration of the log, rather than an end time. • To enter a new end time: a. Press [F2] to edit the end time. b. Press the appropriate feature key to select AM or PM. c.
December 2003 Entering Data Log Parameters (cont.) 8. Entering the Log End Date The next prompt asks you to enter the date you want to end logging. • Press [F1] to accept the displayed date. • Press [F3] to end the log today. • To enter a new date: a. Press [F2] to edit the end date. b. Use the numeric keys to enter the year. Press [ENT]. c. Use the appropriate feature key to select the desired month. Use the [←] [→] keys to view all the options. d. Use the numeric keys to enter the day. Press [ENT].
December 2003 Entering Data Log Parameters (cont.) Once you have completed entering the log parameters, the TransPort returns you to the beginning of the Log Menu. You should now check if there is enough memory available to store the parameters you desire to log as described in the following section. To exit, press the [EXIT] key. The “*” in the indicator box indicates the TransPort is currently logging.
December 2003 Checking the Data Logger Memory Although the TransPort has 120 pages for logger memory, it is a good practice to check the availability of logger memory after setting up a log to see if the amount of data you want to log exceeds the amount of available memory. To check the memory do the following: 1. If you have not, access the Log Menu by pressing the [LOG] key. 2. Press [F2] to select memory.
December 2003 Checking the Data Logger Memory (cont.) Use the following formula to figure out how many pages a log will use: length of log × no. of readings/unit time = no. of records then, no. of records ÷ 120 = no. of pages For example, if you select a log to run for 24 hours with a time increment of 1 minute this log will use 12 pages of memory because: 24 × 60 = 1440 records 1440 ÷ 120 = 12 pages To Stop Logging To stop logging: 1. Press the [LOG] key to access the Log Menu. 2.
December 2003 Error Logging To compile a log of up to three specified error parameters, follow this procedure: 1. Press the [LOG] key to access the Log Menu. 2. Press [F4] to select “ERROR.” 3. Select a name for the log. Use the shift key and the numeric keys to enter the desired name. 4. Enter the log message (if desired) to help identify the particular log. Again, use the shift key and the numeric keys to enter the desired message. 5. The prompt will ask for the first value logged.
December 2003 Chapter 7 Clearing Data On the TransPort, a Clear Menu enables you to delete old logs, site parameter files, and reset totals. To clear parameters, press the [CLR] key. The screen appears similar to that shown in Figure 7-1 below. The indicator box shows you are at the beginning of the Clear Menu. Refer to Appendix D, Menu Maps, Figure D-5 for a flow diagram of the Clear Menu.
December 2003 !WARNING! There is no way to retrieve files once they have been cleared. Be sure you no longer need the information before clearing. Clearing Totals To clear totals: 1. Press [F1] to choose “TOTAL.” 2. Press: • [F1] if you do not want to clear totals. • [F2] to clear totals. The TransPort automatically returns to making measurements. Clearing Site Parameter Files To clear site parameter files: 1. Press [F2] to select “SITE.” 2.
December 2003 Clearing Logs To clear logs: 1. Press [F3] to select “LOG.” 2. Select the log you want to clear by pressing the appropriate feature key. Use the [←] [→] keys to view all the file names. 3. Press: • [F1] if you do not want to clear the log. • [F2] to clear the log. If you have one or more logs left, the TransPort returns to the beginning of the Clear Log option. You may now clear another log or press the [EXIT] key to return to making measurements.
December 2003 Chapter 8 Calibrating Analog Outputs and Inputs To calibrate the analog outputs and inputs, or use the optional ultrasonic thickness gage, you must enter the Calibration Menu by pressing the [CAL] key. When you enter the Calibration Menu the screen appears similar to that in Figure 8-1 below. The Calibration Menu has the three following options (refer to Appendix D, Menu Maps, Figure D-6 for a flow diagram of the Calibration Menu): • A/OUT - enables you to calibrate the analog outputs.
December 2003 Analog Outputs To calibrate the TransPort 4-20 or 0-20 mA output, you need to connect a multimeter or ammeter to the analog output connector shown in Figure 8-2 below (refer to page 2-3 for more information for connecting the analog output). XDCR UP- DOWN- STREAM STREAM PWR 1 ANALOG I/O RS232 1 7 2 3 2 6 FLOW 1 PWR RS232 1. 12VDC-28VDC 6W MAX. 2. RETURN 1. GND(BROWN) 2. TX(RED) 3. RX(ORANGE) 4. DTR(YELLOW) 5. No Connection . (GREEN) 4 6 5 8 ANALOG I/O 1.
December 2003 Analog Outputs (cont.) 4. The screen will display four choices: 4 mA, 20 mA, TEST and EXIT. Press [F1] to select “4 mA”. You must calibrate the low end to 4 mA regardless of the type of output you are using. (If you are using a 0-20 mA output, the TransPort extrapolates the 0 mA point from the 4 mA calibration.) 5. Check the mA reading on the multimeter or ammeter. It should read close to 4 mA. If it does not, use one of the following keys to adjust the reading to 4 mA.
December 2003 Testing the Analog Output Linearity If you select Test in Step 4, the PT868 automatically sets the output to 50%. The resulting mA reading should read close to 12 mA on a 4-20 mA scale (10 mA on a 0-20 mA scale). 1. At the % FULL SCALE prompt, enter a value from 0 to 100 % and press [ENT]. 2. Check the resulting mA reading again. If the reading is off, go back and recalibrate the analog output zero and full scale.
December 2003 Analog Inputs The TransPort enables you to calibrate the analog inputs using a temperature range or a 4 to 20-mA current. To calibrate the temperature input, you need to connect a low and high temperature reference (or calibrated current source) to the temperature input connector shown in Figure 8-3 below. XDCR UP- DOWN- STREAM STREAM PWR 1 ANALOG I/O RS232 1 7 2 3 2 6 FLOW 1 PWR RS232 1. 12VDC-28VDC 6W MAX. 2. RETURN 1. GND(BROWN) 2. TX(RED) 3. RX(ORANGE) 4. DTR(YELLOW) 5.
December 2003 Low and High Calibration Use the following steps to calibrate the inputs: Calibrating the Low Input 1. Connect the low temperature reference (or current source) to the temperature input (see Figure 8-3 on the previous page). 2. If you have not already, enter the Calibration Menu by pressing the [CAL] key. 3. Press [F2] to select “A/IN.” 4. The prompt will ask you to select Input 1 or Input 2. Press [F1] to select Input 1 or [F2] to select Input 2. 5.
December 2003 Low and High Calibration (cont.) You have completed calibrating the analog inputs. You may now do one of the following: • If you are finished calibrating inputs/outputs, press [EXIT] twice and the TransPort begins taking measurements. • If you want to calibrate the analog output, press [EXIT] and refer to page 8-2 for instructions. To reconnect the analog outputs and inputs, refer to Chapter 2, Initial Setup.
December 2003 Chapter 9 Using the Ultrasonic Thickness Gage The TransPort can measure thickness when equipped with the optional ultrasonic thickness transducer (supplied separately). The ultrasonic thickness transducer connects to the transducer connectors. The internal thickness gage not only measures the thickness of a pipe, but can also measure the sound speed of a substance if the thickness is known. The Thickness Gage is factory calibrated and programmed.
December 2003 Connecting the Ultrasonic Thickness Transducer Plug the ultrasonic thickness transducer into the transducer connectors located on the top of the TransPort (see Figure 9-1). Note that the transducer cable plug must be oriented with the center pin down. Thickness Transducer PWR 1 ANALOG I/O RS232 XDCR 1 7 2 3 2 6 4 6 5 8 1 Figure 9-1: Ultrasonic Thickness Transducer Connections !WARNING! When unplugging the transducer, pull on the molded plug only. Do not pull on the cable.
December 2003 Measuring Pipe Thickness The TransPort displays thickness measurements numerically. You may also view the signal and thickness measurements graphically. The graph format displays the signal for as long as you hold the transducer to the pipe. The graph format also has a special feature that enables you to find the signal zero crossing used to calculate the thickness. Use the following sections to make thickness measurements. Making Measurements To measure the pipe wall thickness: 1.
December 2003 Making Measurements (cont.) 3. Use the [←] [→] keys to view all the selections for pipe material. Use the appropriate feature key to make a selection. If your pipe material is not listed, choose the “OTHER” option and enter the longitudinal sound speed of the pipe material. Sound Speeds and Pipe Size Data contains a list of common pipe materials and the corresponding longitudinal sound speeds.
December 2003 Measuring the Sound Speed The Thickness Gage is capable of measuring the sound speed of a material if the thickness of the material is known (at a flange, for example). Use the steps below to measure the sound speed: 1. Press the [CAL] key to enter the Calibration Menu, and then press [F3] to select “WALL”. The screen should appear similar to that shown in Figure 9-4. 2. Press [F3] to select “SNDSP.
December 2003 Graphically Displaying Thickness Transducer Signal To troubleshoot the unit or verify thickness readings, the TransPort enables you to view the thickness transducer signal. To display the thickness gage signal, press the [F1] key to select “GRAPH.” Note: If you are not in the Calibration Menu, press the following keys: [CAL]; [F3] to select “WALL”; [F1] to select “GRAPH.” Figure 9-5 below shows an example of thickness signal.
December 2003 Graphically Displaying Thickness Transducer Signal (cont.) The TransPort displays 120 samples per page. Since there is a limited amount of space to display readings, use the feature key options to view the remaining data. The following options are available (use the [←] [→] keys to view all the options): • [] keys - to move one sample at a time. • [] keys - to pan an entire page. Each page is 3.75 msec.
December 2003 Calibrating the Thickness Gage The ultrasonic thickness gage is factory calibrated; therefore, you only need to perform calibration if you receive new transducers or have a reason to doubt the accuracy of the readings. To calibrate you will need: • a test block, which is shipped with the thickness transducer (or test block of known sound speed and length) • ultrasonic couplant To calibrate the thickness gage, you must enter the Calibration Menu and then enter the appropriate data.
December 2003 Calibrating the Thickness Gage (cont.) 3. The TransPort displays one of the test block lengths. This is the end you should use to calibrate first. Apply couplant to the appropriate end of the test block and place the thickness transducer on the block. The transducer face should be flat against the surface of the test block. 4. The screen should display approximately the correct thickness of the test block. Press the [F1] key to select “CAL.” The TransPort locks in the calibration. 5.
December 2003 Programming the Ultrasonic Thickness Gage Once you access the Thickness Gage option, the TransPort converts from an ultrasonic flowmeter to an ultrasonic thickness gage. The thickness gage has a Program Menu that enables you to view or change the following parameters: • Low Signal Threshold • Detection Threshold • Transducer Delay • Gain Level • Signal Inversion The thickness gage settings are entered at the factory. You should not change them unless instructed by the factory.
December 2003 Programming the Ultrasonic Thickness Gage (cont.) PROGRAM Start LOW SIGNAL THRESHOLD 45 Figure 9-6: Thickness Gage Program Screen Low Signal Threshold This parameter sets the signal strength alarm. If the signal strength falls below this value, a “Low Signal” message appears on the screen. This number should be set to 45 unless otherwise instructed by the factory. To change this number: 1. Use the numeric keys to enter the new number. 2. Press [ENT].
December 2003 Transducer Delay This number represents the time the ultrasonic signal takes to travel from the TransPort to the pipe material surface. This number should be set to 7.0 microseconds unless otherwise instructed by the factory. To change this number: 1. Use the numeric keys to enter the new number. 2. Press [ENT]. The TransPort proceeds to the next prompt.
December 2003 Signal Inversion This prompt enables you to invert the thickness transducer signal. Signal inversion may be required if your thickness readings are showing excessive jitter or for certain unusual combinations of pipe and fluid or pipe and lining. Consult the factory before measuring the thickness of lined pipes.
December 2003 Chapter 10 Diagnostics and Troubleshooting The TransPort is a reliable instrument that is easy to maintain. It will provide accurate flow measurement readings as long as it is operated as described in this manual. If problems do arise with the electronics, transducers or the flowcell, the TransPort displays an error message specifying the possible problem. The TransPort also has a set of diagnostic parameters to help you find and remedy the problem.
December 2003 Error Code Messages Error Code messages appear on the screen while you are displaying measurements. Error Codes do not appear when you are displaying measurements in graph form, or when displaying logged data. The Error Code messages are only general descriptions of the possible problems. Use Table 10-1 on the next page to isolate and remedy the problem. If you are unable to remedy the problem, contact GE Panametrics.
December 2003 Table 10-1: Error Messages Error Message Problem Possible Cause Action E0 No error. Displays briefly after the display of another error message None required. E1 Low Signal Poor ultrasonic signal strength. Broken cable. Flowcell problem. Transducer problem. Electronic failure. Check transducer cable. See Flowcell Problems on page 10-8 and Transducer Problems on page 10-12. Check programming and transducer spacing. Contact the factory if problem persists.
December 2003 Table 10-1: Error Messages (Continued) Error Message 10-4 Problem Possible Cause Action E4 Signal Quality If too high - electronic failure. If too low - flowcell, electrical problem. Check for source of electrical interference. Check electronics with a test flowcell. If unit still fails to operate, contact the factory E5 Amplitude Error Excessive particles or bubbles present in fluid. A lot of second phase present See Flowcell Problems on page 10-8.
December 2003 Displaying Diagnostic Parameters The TransPort provides a Diagnostic Menu to aid in troubleshooting in the event of flowcell, transducer, or electrical problems. Note: The TransPort must be displaying measurements in the Big or Dual Formats as discussed in Chapter 4, Displaying Data, on pages 4-2 and 4-3 to view diagnostic parameters. You can enter the Diagnostic Menu two ways: • Use the right arrow key until “DIAG” appears and then press [F4] to select, or • Press the [EXIT] key.
December 2003 Table 10-2: Diagnostic Parameters Diagnostic Parameter Displays Good Bad SS up Displays the signal strength for the upstream transducer. 50-80 <50 SS do Displays the signal strength for the downstream transducer. 50-80 <50 SNDSP Displays the measured sound speed of the fluid. Check the sound speed of fluid in Sound Speeds and Pipe Size Data. Tup Displays the upstream transit time of the ultrasonic signal in micro seconds.
December 2003 Table 10-2: Diagnostic Parameters (Continued) Diagnostic Parameter Displays Good Bad AMPup Displays the value for the amplitude discriminator of the upstream transducer. 20-28 less than 20 or greater than 28. AMPdn Displays the value for the amplitude discriminator of the downstream transducer. 20-28 fluctuations less than 20 or greater than 28 fluctuations P#up Displays signal peaks for the upstream transducer.
December 2003 Flowcell Problems Flowcell problems fall into two categories: fluid problems and pipe problems. Read the following sections carefully to determine if such problems exist. Fluid Problems Most fluid-related problems can be solved by proper design of the flowmeter installation and the proper choice of transducer frequency. Contact GE Panametrics if you cannot solve a fluidrelated problem. Certain fluid properties may prevent proper flow measurement.
December 2003 Fluid Problems (cont.) 4. THE FLUID SOUND SPEED MUST NOT VARY EXCESSIVELY. The TransPort will tolerate relatively wide, but slow, changes in fluid sound speed due to changes in fluid composition and temperature. However, if you are measuring a fluid that is considerably different from the fluid programmed into the TransPort, you may have to adjust the meter for the new fluid. Refer to Chapter 3, page 3-10, Entering Pipe Parameters.
December 2003 Pipe Problems Improper pipe conditions and/or flowcell installation can cause problems with measurement of the fluid flow. Check the items below if you suspect this class of problem. Contact GE Panametrics if you cannot solve pipe-related problems. 1. THE INSIDE OF THE PIPE OR FLOWCELL MUST BE RELATIVELY CLEAN. Excessive build up of scale, rust, or debris will interfere with flow measurement. Generally, a thin coating or a solid well-adhered buildup on the pipe wall will not cause problems.
December 2003 Pipe Problems (cont.) 4. THE PIPE SURFACES SHOULD BE SMOOTH (FOR CLAMP-ON). When using clamp-on transducers, both the inside and the outside of the pipe at the transducer locations must be smooth. If the pipe is extremely rough, the ultrasonic signal will be scattered by the rough surface, and will not be received by the flowmeter, preventing flow measurement. 5. THE PIPE MUST ALLOW ULTRASOUND TO PASS THROUGH (FOR CLAMP-ON).
December 2003 Transducer Problems Ultrasonic transducers are rugged. However, they are subject to physical damage due to mishandling and chemical attack. Clampon transducers are also subject to installation variables such as physical misalignment and faulty coupling to the pipe on which they are mounted. Because transducer problems are largely dependent on the type of transducer used, wetted or clamp-on, the following list of potential problems is grouped according to transducer type.
December 2003 Wetted Transducer Problems (cont.) 3. INTERNAL DAMAGE An ultrasonic transducer consists of a ceramic “crystal” bonded to the transducer case. The bond between the crystal and the case may be damaged by extreme mechanical shock and by temperature extremes. The crystal itself can also be damaged by the same conditions. The internal wiring can be corroded or shorted if contaminants enter the transducer housing. 4.
December 2003 Clamp-on Transducer Problems 1. POOR COUPLING TO PIPE Clamp-on transducers must be in intimate contact with the pipe. The pipe wall must be smooth and generally free of paint. The couplant material must fill voids between the transducer and the pipe, and must be firmly coupled or bonded to both the pipe and the transducer. The pipe and transducer must be clean and dry for permanent couplant, such as grease or epoxy, to adhere properly.
December 2003 Clamp-on Transducer Problems (cont.) 4. PHYSICAL DAMAGE Transducers may be physically damaged by dropping them onto a hard surface or striking them against another object. Usually the connector on the transducer is the part that is damaged, as it is the most fragile. Minor damage may be repaired by carefully bending the connector back into shape. If the connector cannot be repaired, replace the transducers.
December 2003 Clamp-on Transducer Problems (cont.) If the coupling or pipe wall are not the problem, try relocating the transducers using the following method. However, before beginning, make sure the transducers are not located on or near pipe welds or seams: • Move the upstream transducer about 1/2 inch (12.7 mm) closer to the downstream transducer. If this resolves the problem, you must modify the “S” dimension programmed into the TransPort by using the following steps: 1. Press [PROG] . 2.
December 2003 Chapter 11 Specifications This chapter contains specifications for the following: • Overall • Electrical • Transducer • Fluid Types • Pipe Size and Materials • Available Options Specifications 11-1
December 2003 Overall Flow Accuracy (% of reading): Clamp-on Transducers: Pipe Diameter > 150 mm (6 in.) Velocity >0.3 m/s (>1 ft/s): 2 % typical (1% with calibration) Velocity ≤0.3 m/s (≤1 ft/s): ±0.01 m/s ±0.03 ft/s) typical Pipe Diameter ≤ 150 mm (6 in.) Velocity >0.3 m/s (>1 ft/s): 2-5% typical Velocity <0.3 m/s (<1 ft/s): ±0.05 m/s typical (±0.15 ft/s) Range: 0.03 to 12.2 m/s (0.1 to 40 ft/s) Repeatability: 0.2% to 0.5% clamp-on transducers.
December 2003 Electrical Meter Operating Power: 6 W, maximum. Fuses: One 250 mA, 32V. Two 1 A, 125V, Slo-Blo. Internal Batteries Size: 4 C NiCad high-energy rechargeable batteries. Battery Life: 8-10 hours of continuous operation is typical. The battery life will be reduced if you use the 4-20 mA output and/or loop powered temperature inputs, or the printer output. Recharge Time: 16-20 hours using the internal charger.
December 2003 Operating Modes Flow Measurement: Transit-time clamp-on or wetted transducers Energy: Two loop-powered temperature transmitter inputs to calculate energy flow Input/Output Keypad: 30-key tactile feedback membrane keypad Display: 64 × 128 pixel LCD graphic display Printer/Terminal: One RS232 port for printer, terminal, or PC Analog Output: One 4-20 mA or 0-20 mA current output (550 Ω maximum load):Two outputs for models with serial numbers of 1800 or higher Analog Input: Two 4-20
December 2003 Operational Specifications Site Parameter Programming: Data Logging: Menu-driven operator interface using a keypad and “soft” function keys.On-line help functions including pipe tables. Storage for saving parameters for up to 20 sites. Memory capacity to log up to 43,200 flow data points. Keypad programmable for log units, update times and start and stop times. Display Functions: Graphic displays shows flow in numeric or graphical format. Also displays logged data.
December 2003 Transducer Clamp-On Standard: GE Panametrics’ clamp-on and wetted transducers Standard Temperature Range: -40 to 100° C (-40 to 212° F) Standard Frequency: 1 MHz Note: Transducers for wetted and special applications are available. Flowcells are also available. Consult the factory for details. Optional Materials: Stainless steel, high-temperature plastics, Hastelloy, Monel, Titanium, etc.
December 2003 Pipe Size and Material Clamp-On Transducers: Materials: All metals, most plastics; consult GE Panametrics for concrete, wood, and composite materials for highly corroded, or lined pipes. Pipe Sizes: 12.7 mm to 5 m outside (OD) (1/2 to 200 in.) Pipe Wall Thickness: Up to 75 mm (3 in.) Wetted Transducers: Materials: All metals, most plastics; consult GE Panametrics for concrete, glass, and cement. Pipe Sizes: 3 mm to 5 m inside diameter (ID) (0.125 to 200 in.).
December 2003 Available Options Thickness Gage Standard Transducer: GE Panametrics dual element transducer Pipe Wall Thickness Range: 1.3 to 76.2 mm (0.05 to 3.0 in.) Pipe Materials: Most standard metal and plastic pipe materials Accuracy: 1% typical or ±0.05 mm (±0.002 in.) Display Resolution: Thermal Exposure: 11-8 0.01 mm (0.001 in.) Continuous operation to 37°C (100°F). Intermittent operation to 260° C (500° F) for 10 seconds followed by air cooling for 2 minutes.
December 2003 Appendix A Entering Special Transducer Pipe Parameters If you are using special transducers, the TransPort should be completely programmed when you receive your unit. However, in the unlikely event that data is lost, use this appendix to properly program pipe parameters. The series of prompts that appear in this menu depend on the type of special transducers you are using: wetted or clamp-on.
December 2003 PROGRAM Pipe parameters TRANSDUCER NUMBER 91 Figure A-1: Screen After Pressing PIPE Key 1. Entering the Transducer Number Enter the transducer number. Because you are using a special transducer, do the following: 1. Press [F1] to tell the TransPort you are using a special transducer. 2. Enter a number. The TransPort allows you to use more than one set of special transducers, therefore, assign a number from 91-99. 3. Press [ENT]. 2.
December 2003 Entering Parameters for Clamp-On Transducers If you are using Rayleigh or Shear wave clamp-on transducers, you will get the following prompts: • • • • • • • • • • • • • • • Frequency Tw (time delay) Wedge Angle Wedge Sound Speed Pipe Material Pipe Sound Speed Pipe O.D.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 4C. Entering the Time Delay Tw, the time delay, is actually the time the transducer signal spends travelling through the transducer and cable. This time must be taken into consideration since the TransPort calculates the flow rate by subtracting the time of the upstream and downstream signals. Therefore, the time delay must be subtracted out for an accurate measurement.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 7C. Entering the Pipe Material Enter the pipe material. Since clamp-on transducers send their ultrasonic signals through the pipe wall, the sound speed of the pipe material must be factored in to obtain an accurate flow measurement. Clamp-on transducers are applicable on a variety of pipe materials, only a few of which are among the options for this prompt.
December 2003 Entering Parameters for Clamp-On Transducers (cont.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 8C. Entering the Outside Diameter of the Pipe If you do not know the outside diameter (OD) of your pipe, Sound Speeds and Pipe Size Data contains some standard pipe sizes with their corresponding outside diameters, otherwise measure it. To enter the outside diameter: 1. Press the appropriate feature key to select the units. 2. Use the numeric keys to enter the outside diameter. 3. Press [ENT]. 9C.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 10C. Entering Pipe Lining Clamp-on transducers can be used on pipe with a variety of pipe linings; however, the options list only a few types. If your lining is not among the options, select “OTHER” and enter the sound speed of your lining material from the tables in Sound Speeds and Pipe Size Data. If your pipe does not have a lining, press [F1] to respond no and the TransPort proceeds to the next prompt.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 11C. Selecting Tracking Windows The next prompt asks if you want tracking windows. • Press [F1] for no or • Press [F2] for yes. 12C. Entering the Fluid Type The next prompt asks you to enter the fluid type.
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 13C. Entering the Reynolds Correction Factor The Reynolds Correction Factor is a number based on the viscosity and flow rate of the fluid. Generally, it should only be enabled when using transducers mounted in a diametrical path (such as clamp-on or tilted diameter). If you enable the Reynolds Correction Factor, you will also need to enter the Kinematic Viscosity of the fluid (units are ft2/sec or m2/sec depending on the units selected).
December 2003 Entering Parameters for Clamp-On Transducers (cont.) 16C. Display of Transducer Spacing The next prompt shows the suggested spacing of the transducers calculated from the information entered. Use the Liquid Transducer Installation Guide as a guide to installing the transducers on the pipe with the spacing dimension (S) given here. Set the transducers to the spacing shown at this prompt.
December 2003 Entering Parameters for Wetted Transducers If you are using wetted transducers, you will see the following prompts: • Frequency • Tw (time delay) • Pipe OD (pipe outside diameter) • Pipe Wall • Path Length (P) • Axial Dimension (L) • Tracking Windows • Fluid Type • Reynolds Correction Factor • Calibration Factor 3W. Entering the Frequency Enter the transducer frequency.
December 2003 Entering Parameters for Wetted Transducers (cont.) 4W. Entering the Time Delay Tw, the time delay, is actually the time the transducer signal spends travelling through the transducer and cable. This time must be taken into consideration since the TransPort calculates the flow rate by subtracting the time of the upstream and downstream signals. Therefore, the time delay must be subtracted out for an accurate measurement. Use the numeric keys to enter the time delay (in microseconds).
December 2003 Entering Parameters for Wetted Transducers (cont.) 6W. Entering the Pipe Wall Thickness If you are not sure of the pipe wall thickness, Sound Speeds and Pipe Size Data contains some standard pipe sizes with their corresponding wall thicknesses. Also, the pipe wall thickness can be determined by using the optional internal ultrasonic thickness gage as described in Chapter 9, Using the Ultrasonic Thickness Gage. Use the numeric keys to enter the pipe wall thickness. Press [ENT]. 7W.
December 2003 Entering Parameters for Wetted Transducers (cont.) 8W. Enter the Axial Length (L) GE Panametrics has calculated the axial length (L) of the ultrasonic signal based on the transducer configuration for your particular application. Find the axial dimension on the flowcell or from other supplied documentation. 1. Press the appropriate feature key to select the desired units. 2. Use the numeric keys to enter the axial length. Press [ENT].
December 2003 Entering Parameters for Wetted Transducers (cont.) If you have enabled the tracking windows,you may choose one of the following: • Water, 0-100° C • Water, 0-260° C • Oil • Other — Enter sound speed (c3) limits Use the [←] [→] keys to view all the available options. Then use the appropriate feature key to make a choice. Depending on the fluid type you choose, the TransPort may ask you for additional information such as fluid temperature or sound speed. 11W.
December 2003 Entering Parameters for Wetted Transducers (cont.) You have completed entering data into the PIPE option. You may at this time do one of the following: • Use the [↑] and [↓] arrow keys to verify or change data. • Continue to enter data in other menus as described in Chapter 3, Programming Site Data. • Press [EXIT] to exit the Program Menu and begin taking measurements. If you decide to exit, you must have already programmed the system parameters in order to make flow measurements.
December 2003 Appendix B Calculating P and L Dimensions If you are using wetted transducers, the TransPort requires that you enter the path length and the axial dimension. These dimensions are referred to as P and L. P is the transducer face to face distance and L is the axial projection of P in the flow stream. To determine L, physically measure the distance between the center of the transducer ports at the inside wall as shown in Figure B-1, if possible. If not, consult the factory.
December 2003 To determine P, you need the following: • the pipe inside diameter (ID) • the wall thickness (WT) • the installed pipe coupling length (CL) • the transducer face depth (FD) • the mounting angle (MA) Use Figure B-2 to properly measure the coupling length. Typically, the transducer face is positioned just outside the inside diameter (ID) of the pipe, or slightly retracted inside the coupling.
December 2003 Use the following formula to determine the P dimension: [ID + 2(WT)]/(cos MA) + 2(CL - FD) = P Dimension For example, given the following: • a stack inside diameter of 48 in. • wall thickness of 3/8 in. • installed coupling length of 2.0 in. • a transducer face depth of 1.75 in. • 45 degree mounting angle The P dimension would be [48 in. + 2(3/8 in.)]/(0.7071) + 2(2.0 in. - 1.75 in.) = 69.4 in.
December 2003 Appendix C Serial Connections The TransPort can transmit stored data and displayed readings using the RS232 communication port. In order to enable the TransPort to communicate properly, you must use the proper cable. Panametrics provides the following cables which are compatible with most computer or serial devices (see Table C-1). Table C-1: Serial Port Cables Cable Part # Connector Description Cable Length Typical Device 26DLC/9F-6 “D” Type, 9 pin female contacts 1.8 m (6 ft) 3.
December 2003 If you wish to construct a suitable cable, use the information in Table C-2 to make the RS232 pin connections.
December 2003 Appendix D Menu Maps This appendix contains menu maps for each of the TransPort functions, which are as follows: • PROG - enter site location data into the Program Menu. • PRINT - print stored and displayed data. • LOG - log data. • DISP - set up the screen to display measurements in text or graph form. • CLR - clear log, site and totals. • CAL - calibrate the analog inputs and outputs, as well as use the optional ultrasonic thickness gage.
December 2003 Program Menu ACTIV Refer to D-1A SYSTM Refer to D-1B PIPE Refer to D-1C and D-1D I/O Refer to D-1E, D-1F, D-1G and D-1H PROG SAVE Refer to D-1I RECLL Refer to D-1J COMM Refer to D-1K SETUP Refer to D-1L Figure D-1: PROGRAM Menu D-2 Menu Maps
December 2003 PROG ACTIV Trans EXIT Figure D-1A: ACTIV Option in the PROGRAM Menu Menu Maps D-3
December 2003 PROG SYSTM Energy Option Site Message System Units Stopwatch Totalizer Volumetric Units Totalizer Units If the Energy Option is ON Power Energy (Total) If the Energy Option is OFF Heat or Cool Sys Flow Measurement Date Time EXIT Figure D-1B: SYSTeM Option in the PROGRAM Menu D-4 Menu Maps
December 2003 PROG PIPE Transducer # Wedge Temp Input (or med + high temp) Active or Fixed If you are using clamp-on or wetted Pipe Material Pipe O.D. Pipe Wall If you are using clamp-on or wetted Lining Path Length P Lining Material Axial Length L Lining Thickness Tracking Windows Fluid Type Reynolds Correction Kin. Viscosity Only appears if Reynolds Correction is enabled If you are using clamp-on or wetted Kin. Viscosity Cal.
December 2003 PROG Transducer # PIPE Press [F1] to select SPEC. Special Transducer # Wedge Type Frequency Tw If you are using clamp-on or wetted Wedge Angle Wedge Sound Speed Pipe Material Pipe Sound Speed Pipe OD Pipe Wall Lining If you are using clamp-on or wetted Path Length P Lining Material Axial Length L Lining Thickness Tracking Windows Fluid Type Reynolds Correction Kin.
December 2003 PROG I/O Error Handling IN1 IN2 OUT1 OUT2 Figure D-1E: Input/Output Option in the PROGRAM Menu Menu Maps D-7
December 2003 PROG I/O Error Handling Hold Low High Zero Cutoff EXIT Figure D-1F: Error Handling in the Input/Output Option D-8 Menu Maps
December 2003 PROG I/O IN1 If Energy Option Off: IN2 If Energy Option On: If Energy Option Off: If Energy Option On: Temp Input Return Temp Input Supply Fixed or Active Fixed or Active Return Base Supply Base Supply Full Return Full Return Temperature Supply Temperature Analog Input 2 Analog Input 1 Off Off Other Other Input Name Input Name Input Units Input Units Base (4 mA) Base (4 mA) Full (20 mA) Full (20 mA) EXIT Figure D-1G: Inputs in the Input/Output Option Menu Maps D-9
December 2003 PROG I/O OUT1 Off OUT2 0-20 mA 4-20 mA Off Output Measurement 0-20 mA 4-20 mA Output Measurement Base Base Full Full EXIT Figure D-1H: Outputs in the Input/Output Option D-10 Menu Maps
December 2003 PROG SAVE Name Enter a new name and press [ENT], or use a feature key to select a file to overwrite.
December 2003 PROG RECLL Name EXIT Figure D-1J: RECaLL Option in the PROGRAM Menu D-12 Menu Maps
December 2003 PROG COMM Baud Rate UART bits Network ID? EXIT Figure D-1K: COMMunication Option in the PROGRAM Menu Menu Maps D-13
December 2003 PROG SETUP SIGNL Signal Low Limit Signal High Limit Cor. Peak Limit Soundspeed +-Limit Velocity Low Limit Velocity High Limit Acceleration Limit INIT DIGIT AVRG Response Time Vol. Decimal Digits Total Decimal Digits Power Decimal Digits Energy Decimal Digits For Energy Option Amp. Discrim. Low Amp. Discrim. High Delta T Offset % of Peak Transmitter Voltage Xmit Sample Size # Errors Allowed Min. Peak % Limit Max.
December 2003 Print Menu PRNT DATA Refer to D-2A LOG Refer to D-2B PROG Refer to D-2C STOP Refer to D-2D PRNTR Refer to D-2E Figure D-2: PRiNT Menu Menu Maps D-15
December 2003 PRNT DATA Time Increment Format Num or Plot Graph Variable Y Axis Max. Y Range Time Increment The TransPort sends information to the printer and returns to making measurements.
December 2003 PRNT LOG Format Num or Plot Name Name Y Axis Max. Y Range The TransPort sends the log to the printer and returns to making measurements.
December 2003 PRNT PROG Name The TransPort sends the selected site location file to the printer and returns to making measurements.
December 2003 PRNT STOP Stop Printing No or Yes The TransPort continues printing the current file and returns to making measurements. The TransPort stops printing the current file and returns to making measurements.
December 2003 PRNT PRNTR Printer Type The TransPort returns to making measurements.
December 2003 LOG STOP STD MEM STOP ERROR Name X/X Pages FREE X Pages PENDING Name Name Log Message Press [ENT] 1st Value Logged 2nd Value Logged The TransPort continues logging. 3rd Value Logged Now The TransPort stops logging the selected file.
December 2003 BIG Display Menu DUAL GRAPH Graph Variable Time Increment Y Axis Max. Y Range DISP LOG Name NUM PLOT Y Axis Max. Y Range SIGNL BACKL Backlight Timeout The new display appears on the TransPort screen.
December 2003 CLR TOTAL SITE LOG Clear Total? Name Name No or Yes Clear Site? Clear Log Name? No or Yes No or Yes The TransPort clears the totals and returns to making measurements. The TransPort returns to making measurements. The TransPort clears the selected site. The TransPort clears the log and returns to making measurements.
December 2003 CAL A/IN A/OUT WALL Graph Mat’l Sndsp Output 1 Output 2 4 mA 20 mA Up Exit Test Connect temp. input Down Input 1 Store Input 2 The TransPort displays the soundspeed. Abort EXIT Low High 4 mA 20 mA EXIT Low Reference Temp The TransPort displays the thickness transducer signal. High Reference Temp EXIT The TransPort displays the sample thickness.
December 2003 CAL WALL PROG Low Signal Threshold CAL Detection Threshold Block #1 Length Transducer Delay CAL EXIT The PT868 calibrates block #1. System Units Signal Inversion Gain Level Block #2 Length EXIT CAL The PT868 calibrates block #2.
December 2003 Appendix E Temperature Transmitter Installation The TransPort requires a supply and return temperature input to measure energy rate and consumption. You must connect the temperature sensors to a 4 to 20 mA transmitter (powered by the TransPort or externally) and then from the transmitter to the TransPort (a special GE Panametrics cable is required to make transmitter to TransPort connections).
December 2003 Guidelines for RTD Installation You should have discussed possible installation sites with an applications engineer or field sales person by the time you receive the RTDs. However, you should follow these guidelines to ensure accurate temperature measurement: • Do not install the RTD on a pipe seam. • Clean the pipe at the installation location. • Do not install the RTD on the top or the bottom of the pipe. • Apply thermal couplant to the RTD before mounting it to the pipe.
December 2003 Assembling the Clamping Fixture (cont.) 5. Each RTD has a securing post (located on the rounded portion of the RTD) that secures the RTD in place when it is mounted on the pipe. Place the RTD on the strap by sliding the strap under the securing post. IMPORTANT: Make sure you place the RTD on the strap so when the strap is wrapped around the pipe, the flat surface of the RTD sits against the pipe. 6.
December 2003 Making Electrical Connections The TransPort will not accept a signal directly from the RTD; therefore, you must have some type of 4 to 20-mA transmitter. GE Panametrics supplies a dual transmitter (DTR) with a special LEMO connector that attaches to the TransPort. The TransPort supplies power to the DTR using an internal 16 V supply.
December 2003 Connecting the Transmitter to the TransPort If you are using the GE Panametrics DTR, simply plug the LEMO connector into the ANALOG I/O connector as shown in Figure E-1. XDCR UP- DOWN- STREAM STREAM PWR 1 ANALOG I/O RS232 1 7 2 3 2 6 FLOW 1 PWR RS232 1. 12VDC-28VDC 6W MAX. 2. RETURN 1. GND(BROWN) 2. TX(RED) 3. RX(ORANGE) 4. DTR(YELLOW) 5. No Connection . (GREEN) 4 6 5 8 ANALOG I/O 1. 4-20mA/0-10V ANALOG OUT 2. 16V (switched) 3. SUPPLY TEMP 4. RETURN TEMP 5. GND 6. AOUT2 7.
December 2003 Connecting the Transmitter to the TransPort (cont.) If you are using your own transmitter, you must connect the special GE Panametrics cable to your transmitter as shown in Table E-1.
December 2003 Appendix F Ultrasonic Thickness Gage Theory of Operation All ultrasonic thickness gaging involves timing the round trip of a sound pulse in a test material. Because solid metal has an acoustic impedance that differs from that of gasses, liquids, or corrosion products such as scale or rust, the sound pulse will reflect from the far surface of the remaining metal.
December 2003 Factors Affecting Performance and Accuracy A. Surface Condition - Loose or flaking scale, rust, corrosion or dirt on the outside surface of a test piece will interfere with the coupling of sound energy from the transducer into the test material. Thus, any loose debris of this sort should be cleaned from the specimen with a wire brush or file before measurements are attempted.
December 2003 Factors Affecting Performance and Accuracy (cont.) As with any difficult application, experimentation with actual product samples is the best way to determine the limits of a particular gage/transducer combination on a given surface. B. Transducer Positioning/Alignment - For proper sound coupling the transducer must be pressed firmly against the test surface.
December 2003 Factors Affecting Performance and Accuracy (cont.) Remember that an ultrasonic test measures thickness at only one point within the beam of the transducer, and that in corrosion situations wall thicknesses often vary considerably. Test procedures usually call for making a number of measurements within a defined area and establishing a minimum and/or average thickness.
December 2003 Factors Affecting Performance and Accuracy (cont.) D. Taper or Eccentricity - If the contact surface and the back surface are tapered or eccentric with respect to each other, the return echo again becomes distorted and the accuracy of measurement is diminished. E. Acoustic Properties of the Material - There are several conditions found in engineering materials that can severely limit the accuracy and thickness range that can be measured. 1.
December 2003 Appendix G Screen and Error Messages This appendix contains the common screen and error messages that may appear on the TransPort screen either while the unit is making measurements or while you are performing a task. The table below lists the messages alphabetically. If the message begins with a number, use the second word in the message when looking it up. Symbols are listed first in the table in random order. Message Description/Action * Indicates the TransPort is logging data.
December 2003 Message Description/Action All Sites Cleared! hit key This message may appear while performing the following: •saving a new site - the user did not enter a new name or select to overwrite an existing site; •recalling a site - there are no site files to recall; •clearing a site - there are no sites to delete; •printing a site - there are no sites to print; •displaying a site - there are no sites to display. Hit any key and the TransPort returns to making measurements.
December 2003 Message Description/Action E1: Low Signal This message appears on the message line on the screen. This indicates that the ultrasonic signal is poor and exceeds the limits programmed in the PROGRAM Menu SETUP option. Check the transducer as described on page 10-12. E2: Sound Speed Error This message appears on the message line on the screen. This indicates that the sound speed for your fluid exceeds the limits programmed in the PROGRAM Menu SETUP option.
December 2003 Message Description/Action E6: Cycle Skip, Accel This message appears on the message line on the screen. This indicates that the acceleration exceeds the limits programmed in the PROGRAM Menu SETUP option. Refer to page 10-4. E7: Analog Out Error This message appears on the message line on the screen. Current or voltage is over the limit. Refer to page 10-4. E8: TEMPerature INput SUPPLY This message appears on the message line on the screen. Supply temperature input is out of range.
December 2003 Message Description/Action Log Active, END only hit any key This message appears when in the LOG Menu. This indicates the Log is still compiling data. You may only edit the End Time Log DONE, to inspect hit any key This message appears when in the LOG Menu. This indicates the Log is complete. Hit any key and the TransPort displays the log on the screen. Low Signal This message appears while using the thickness gage. This indicates the input device signal is poor.
December 2003 Message S L Description/Action This message appears in the indicator box. SL indicates the shift key function is locked. Press the shift key to unlock. Start This message appears in the indicator box. This message appears frequently when performing all tasks. Start indicates that you are at the beginning of a task menu or option. Starting time must exceed current time! This message appears while in the LOG menu STD option.
December 2003 Appendix H RTD Temperature vs. Resistance Table Table H-1: RTD Temperature Vs. Resistance For European Curve, Alpha = 0.00385 1° Celsius Increments °C Ohm -200 18.49 -199 18.93 -198 Diff. °C Ohm Diff. -184 25.37 0.43 0.44 -183 25.80 0.43 19.36 0.43 -182 26.23 0.43 -197 19.79 0.43 -181 26.65 0.42 -196 20.22 0.43 -180 27.08 0.43 -195 20.65 0.43 -179 27.50 0.42 -194 21.08 0.43 -178 27.93 0.43 -193 21.51 0.43 -177 28.35 0.42 -192 21.94 0.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. -167 32.59 0.42 -145 41.79 0.41 -166 33.01 0.42 -144 42.21 0.42 -165 33.43 0.42 -143 42.63 0.42 -164 33.85 0.42 -142 43.04 0.41 -163 34.27 0.42 -141 43.45 0.41 -162 34.69 0.42 -140 43.87 0.42 -161 35.11 0.42 -139 44.28 0.41 -160 35.53 0.42 -138 44.70 0.42 -159 35.95 0.42 -137 45.11 0.41 -158 36.37 0.42 -136 45.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. -123 50.88 0.41 -101 59.85 0.41 -122 51.29 0.41 -100 60.25 0.40 -121 51.70 0.41 -99 60.66 0.41 -120 52.11 0.41 -98 61.06 0.40 -119 52.52 0.41 -97 61.47 0.41 -118 52.92 0.40 -96 61.87 0.40 -117 53.33 0.41 -95 62.28 0.41 -116 53.74 0.41 -94 62.68 0.40 -115 54.15 0.41 -93 63.09 0.41 -114 54.56 0.41 -92 63.49 0.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. -79 68.73 0.40 -57 77.52 0.39 -78 69.13 0.40 -56 77.92 0.40 -77 69.53 0.40 -55 78.32 0.40 -76 69.93 0.40 -54 78.72 0.40 -75 70.33 0.40 -53 79.11 0.39 -74 70.73 0.40 -52 79.51 0.40 -73 71.13 0.40 -51 79.91 0.40 -72 71.53 0.40 -50 80.31 0.40 -71 71.93 0.40 -49 80.70 0.39 -70 72.33 0.40 -48 81.10 0.40 -69 72.73 0.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. -35 86.25 0.40 -13 94.91 0.39 -34 86.64 0.39 -12 95.30 0.39 -33 87.04 0.40 -11 95.69 0.39 -32 87.43 0.39 -10 96.09 0.40 -31 87.83 0.40 -9 96.48 0.39 -30 88.22 0.39 -8 96.87 0.39 -29 88.62 0.40 -7 97.26 0.39 -28 89.01 0.39 -6 97.65 0.39 -27 89.40 0.39 -5 98.04 0.39 -26 89.80 0.40 -4 98.44 0.40 -25 90.19 0.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 9 103.51 0.39 31 112.06 0.39 10 103.90 0.39 32 112.45 0.39 11 104.29 0.39 33 112.83 0.38 12 104.68 0.39 34 113.22 0.39 13 105.07 0.39 35 113.61 0.39 14 105.46 0.39 36 113.99 0.38 15 105.85 0.39 37 114.38 0.39 16 106.24 0.39 38 114.77 0.39 17 106.63 0.39 39 115.15 0.38 18 107.02 0.39 40 115.54 0.39 19 107.40 0.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 53 120.55 0.39 75 128.98 0.38 54 120.93 0.38 76 129.37 0.39 55 121.32 0.39 77 129.75 0.38 56 121.70 0.38 78 130.13 0.38 57 122.09 0.39 79 130.51 0.38 58 122.47 0.38 80 130.89 0.38 59 122.86 0.39 81 131.27 0.38 60 123.24 0.38 82 131.66 0.39 61 123.62 0.38 83 132.04 0.38 62 124.01 0.39 84 132.42 0.38 63 124.39 0.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 97 137.36 0.38 119 145.68 0.37 98 137.74 0.38 120 146.06 0.38 99 138.12 0.38 121 146.44 0.38 100 138.50 0.38 122 146.81 0.37 101 138.88 0.38 123 147.19 0.38 102 139.26 0.38 124 147.57 0.38 103 139.64 0.38 125 147.94 0.37 104 140.02 0.38 126 148.32 0.38 105 140.39 0.37 127 148.70 0.38 106 140.77 0.38 128 149.07 0.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 141 153.95 0.37 163 162.16 0.37 142 154.32 0.37 164 162.53 0.37 143 154.70 0.38 165 162.90 0.37 144 155.07 0.37 166 163.27 0.37 145 155.45 0.38 167 163.65 0.38 146 155.82 0.37 168 164.02 0.37 147 156.19 0.37 169 164.39 0.37 148 156.57 0.38 170 164.76 0.37 149 156.94 0.37 171 165.13 0.37 150 157.31 0.37 172 165.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 185 170.31 0.37 207 178.41 0.37 186 170.68 0.37 208 178.78 0.37 187 171.05 0.37 209 179.14 0.36 188 171.42 0.37 210 179.51 0.37 189 171.79 0.37 211 179.88 0.37 190 172.16 0.37 212 180.24 0.36 191 172.53 0.37 213 180.61 0.37 192 172.90 0.37 214 180.97 0.36 193 173.26 0.36 215 181.34 0.37 194 173.63 0.37 216 181.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 229 186.45 0.36 251 194.44 0.37 230 186.82 0.37 252 194.80 0.36 231 187.18 0.36 253 195.16 0.36 232 187.54 0.36 254 195.52 0.36 233 187.91 0.37 255 195.88 0.36 234 188.27 0.36 256 196.24 0.36 235 188.63 0.36 257 196.60 0.36 236 189.00 0.37 258 196.96 0.36 237 189.36 0.36 259 197.33 0.37 238 189.72 0.36 260 197.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 273 202.36 0.35 295 210.24 0.36 274 202.72 0.36 296 210.59 0.35 275 203.08 0.36 297 210.95 0.36 276 203.44 0.36 298 211.31 0.36 277 203.80 0.36 299 211.66 0.35 278 204.16 0.36 300 212.02 0.36 279 204.52 0.36 301 212.37 0.35 280 204.88 0.36 302 212.73 0.36 281 205.23 0.35 303 213.09 0.36 282 205.59 0.36 304 213.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 317 218.05 0.35 339 225.81 0.35 318 218.41 0.36 340 226.17 0.36 319 218.76 0.35 341 226.52 0.35 320 219.12 0.36 342 226.87 0.35 321 219.47 0.35 343 227.22 0.35 322 219.82 0.35 344 227.57 0.35 323 220.18 0.36 345 227.92 0.35 324 220.53 0.35 346 228.27 0.35 325 220.88 0.35 347 228.62 0.35 326 221.24 0.36 348 228.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 361 233.52 0.35 383 241.17 0.35 362 233.87 0.35 384 241.51 0.34 363 234.22 0.35 385 241.86 0.35 364 234.56 0.34 386 242.20 0.34 365 234.91 0.35 387 242.55 0.35 366 235.26 0.35 388 242.90 0.35 367 235.61 0.35 389 243.24 0.34 368 235.96 0.35 390 243.59 0.35 369 236.91 0.35 391 243.93 0.34 370 236.65 0.34 392 244.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 405 248.76 0.35 427 256.29 0.34 406 249.10 0.34 428 256.64 0.35 407 249.45 0.35 429 256.98 0.34 408 249.79 0.34 430 257.32 0.34 409 250.13 0.34 431 257.66 0.34 410 250.48 0.35 432 258.00 0.34 411 250.82 0.34 433 258.34 0.34 412 251.16 0.34 434 258.68 0.34 413 251.50 0.34 435 259.02 0.34 414 251.85 0.35 436 259.
December 2003 Table H-1: RTD Temperature Vs. Resistance (Continued) For European Curve, Alpha = 0.00385 °C Ohm Diff. °C Ohm Diff. 449 263.87 0.34 465 269.18 0.34 450 264.11 0.34 466 269.51 0.33 451 264.45 0.34 467 269.85 0.34 452 264.79 0.34 468 270.19 0.34 453 265.13 0.34 469 270.52 0.33 454 265.47 0.34 470 270.86 0.34 455 265.80 0.33 471 271.20 0.34 456 266.14 0.34 472 271.53 0.33 457 266.48 0.34 473 271.87 0.34 458 266.82 0.34 474 272.
December 2003 Appendix I Material Safety Data Sheet for Couplant Material Safety Data Sheet for Couplant I-1
December 2003 Material Safety Data Sheet (To comply with 29 CFR 1910.1200) Effective Date 4/1/98 Note: N/A = not applicable or not available Section 1— Product Identification Product Name: SOUNDSAFE® Generic Name: Ultrasonic Couplant Manufacturer: Sonotech, Inc. 774 Marine Dr., Bellingham, WA 98225 360-671-9121 FAX: 360-671-9024 E-mail: sonotech@nas.com http://www.sonotech-inc.com NFPA Hazardous Materials Identification System (est) Health...................................0 Flammability .................
December 2003 Section 3 — Physical Data (nominal) Boiling Point: >220°F Freezing Point: <20°F Vapor Pressure: N/A Evaporation Rate: N/A Solubility in Water: complete Appearance and Odor: water white, opaque gel; bland odor pH: 7.35 — 7.9 Acoustic Imp.: 1.726 X 106 Vapor Density: N/A Specific Gravity: 1.
December 2003 Section 6 — Health Hazard and First Aid Data Routes of Entry: 1 Skin: not likely Eyes: not normally Ingestion: not normally Inhalation: no 1 SOUNDSAFE® contains only food grade and cosmetic grade ingredients. Effects of Overexposure: Acute: May cause temporary eye irritation. Chronic: None expected. First Aid Procedures: Skin: Remove with water if desired. Eyes: Flush with water for 15 minutes. Ingestion: For large quantities, induce vomiting and call a physician.
December 2003 Section 8 — Control Measures Respiratory Protection: not required Ventilation: not required Protective Gloves: on individuals demonstrating sensitivity to SOUNDSAFE® Eye Protection: as required by working conditions Other Protective Equipment: not required Material Safety Data Sheet for Couplant I-5
December 2003 Index A Acceleration Limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-42 ACTIV Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 Adjustments Screen Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 Amplitude Discriminator Enter High Limit . . . . . . . . . . . . . . . . .
December 2003 Index (cont.) C Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1 Analog Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 Temperature Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 Thickness Gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8 Calibration Factor Enter . . . . . . . . . . . . . . . . . . . . . . . . . . .
December 2003 Index (cont.) D Date Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10 Decimal Digits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-46 Description Electronics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2 Flowcell. . . . . . . . . . . . . . . . . . . . . . .
December 2003 Index (cont.) F Feature Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11 Flowcell Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Fluid Problems Flowcell Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8 Fluid Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19 Enter . . . . . . . . . . . . . . . . . . . . . .
December 2003 Index (cont.) L L Dimension Calculate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1 Limits Signal Limits Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40 Lining Enter Pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18 Log End Start Date. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6 Log Format . . . . . . . . . . . . . . . . .
December 2003 Index (cont.) O On-Line Help Getting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 Outputs Analog Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Outside Diameter Enter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17, 3-22 P P and L Dimensions . . . . . . . . . .
December 2003 Index (cont.) Problems Flowcell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8 Fluid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8 Pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-10 Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-12 Program Menu . . . . . . . . . . . . . . . . . . . . . . . .
December 2003 Index (cont.) R Recall Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34 Resetting Setup Parameters Initializing Setup Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-46 Response Time Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-46 Retrieve Site Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34 RS232 Communication. . . . . . . . . . .
December 2003 Index (cont.) Sound Speed Limit Enter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-41 Specifications Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3 Overall. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2 Pipe Size and Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7 Thickness Gage . . . . . . . . . . . . . .
December 2003 Index (cont.) Connecting Thickness Gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2 Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Display Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 Enter Clamp-On Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Enter Parameters for Special Transducers . . . . . . . . . . . . . . . . . . A-1 Enter Spacing . . . .
DECLARATION OF CONFORMITY GE Panametrics Shannon Industrial Estate Shannon, Co.
DECLARATION DE CONFORMITE GE Panametrics Shannon Industrial Estate Shannon, Co.
KONFORMITÄTSERKLÄRUNG GE Panametrics Shannon Industrial Estate Shannon, Co.
WORLDWIDE OFFICES MAIN OFFICES: GE PANAMETRICS INTERNATIONAL OFFICES: USA GE Panametrics 221 Crescent St., Suite 1 Waltham, MA 02453-3497 USA Telephone: 781-899-2719 Toll-Free: 800-833-9438 Fax: 781-894-8582 E-mail: panametrics@ps.ge.com Web: www.gepower.com/panametrics ISO 9001 Certified Australia P.O. Box 234 Gymea N.S.W. 2227 Australia Telephone 61 (02) 9525 4055 Fax 61 (02) 9526 2776 E-mail panametrics@panametrics.com.au Japan 2F, Sumitomo Bldg.
USA GE Panametrics 221 Crescent Street, Suite 1 Waltham, MA 02453-3497 Telephone: (781) 899-2719 Toll-free: (800) 833-9438 Fax: (781) 894-8582 E-Mail: panametrics@ps.ge.com Web: www.gepower.com/panametrics Ireland GE Panametrics Shannon Industrial Estate Shannon, County Clare Ireland Telephone: 353-61-470200 Fax: 353-61-471359 E-Mail: info@panametrics.