Specifications
23
START UP SERVICE
Environmental Monitor Service, Inc. is available to assist you, call our service department at (203) 935-0102 or fax
(203) 634-6663 for details and to schedule start up.
Verifying instrument operation and configuration
Alignment procedure must be completed as outlined in Section 2.
Power applied to both the control unit and the sensors for a minimum of 30 Minutes before any adjustments are
attempted.
Observe the front panel LED's and confirm the red Operate and the green RUN LED's are on. Press the ACK
button to confirm NO ALM’S message is displayed. Press the DIS button until OPACITY % is displayed.
EDIT KEYS
All data entry, editing, and diagnostics are accomplished with four EDIT
keys behind the hinged cover.
FACTORY O.P.L.R. DISPLAY
Under the flip down door, press the EDIT (ESC) button (the red Edit
(ESC) LED will light). The message display will read QUICK MENU. Press
the ENTER key, then the EDIT key or hold and scroll down until the
Display reads “OPLR Read Only”. Press the ENTER key the OPLR is
displayed. This is a read only block and cannot change the OPLR of the
system. It is intended for a convenient place to record the OPLR. The OPLR
is factory set and cannot be changed in the field. If an OPLR change is
required contact the factory for return authorization number and shipping
instructions.
QUICK MENU
Alarm Set Point In % opacity
Alarm Time Delay In Seconds
Auto Cal Cycle Time In hours
Zero Cal Value In % opacity
Span Cal Value In % opacity
OPLR Read Only XX.XX
(ESC)
EDIT
ENTER
EDIT
EDIT
REM
LOC
AUTO
MAN
DIS
ACK
+19999
MESSAGE
PV
SP
OUT
EDIT
Keys
Hinged
Cover
“Edit
Value”
Light
“Scroll
Menu”
light
Moves
Message
Cursor