Satellite Radio User Manual
19
If using shielded or armored cable:
a. Ground the shield drain wires (the black wire) only on the core 
processor end, by connecting it to the ground screw inside the 
lower conduit ring. Do not ground to the core processor’s 
mounting screw. Do not ground the cable at the sensor junction 
box. 
b. Ground the cable braid on both ends, by terminating it inside the 
cable glands.
4. Ensure integrity of gaskets, grease all O-rings, then close the junction 
box housing and core processor end-cap, and tighten all screws.
STEP 8. Connecting power supply wiring
CAUTION
Damaging the wires that connect the transmitter to 
the sensor can cause measurement error or meter 
failure.
To reduce the risk of measurement error or meter failure, 
when closing the housings on the sensor and core 
processor, make sure that the wires are not caught or 
pinched.
CAUTION
Improper wiring installation can cause device failure 
or measurement error.
• To avoid device failure or measurement error, do not 
install power supply wiring in the same cable tray or 
conduit as input/output wiring.
• Shut off power supply before installing the applications 
platform.
• Make sure power supply voltage matches voltage that 
is indicated on power supply wiring terminals. See 
Figure 12.