Satellite Radio User Manual
14
Wiring instructions for remote core processor with remote transmitter
There are two phases to this procedure: 
• Wiring the remote core processor to the transmitter
• Wiring the sensor to the remote core processor
To wire the remote core processor to the transmitter:
1. Use one of the following methods to shield the wiring:
• If you are installing unshielded wiring in continuous metallic 
conduit that provides 360° termination shielding for the enclosed 
wiring, go to Step 6.
• If you are installing a user-supplied cable gland with shielded 
cable or armored cable, terminate the shields in the cable gland. 
Terminate both the armored braid and the shield drain wires in the 
cable gland. Go to Step 6.
• If you are installing a Micro Motion-supplied cable gland at the 
core processor housing:
- If you are using shielded cable, prepare the cable and apply 
shielded heat shrink as described in Step 4. The shielded heat 
shrink provides a shield termination suitable for use in the 
gland when using cable whose shield consists of foil and not 
a braid. Go to Step 2.
- If you are using armored cable, prepare the cable as described 
in Step 4, but do not apply heat shrink – omit Steps 4d, e, f, 
and g. Go to Step 2.
2. Identify the components shown in Figure 10. Remove the core 
processor lid.
3. Slide the gland nut and the clamping insert over the cable.










