Instruction manual
MODEL XMT-A-FF/FI SECTION 13.0
CALIBRATION - OZONE
13.3 PROCEDURE — FULL SCALE CALIBRATION
1. Place the sensor in the process liquid. Adjust the sample flow until it is within the range recommended for the
sensor. Refer to the sensor instruction sheet.
2. Adjust the ozone concentration until it is near the upper end of the operating range. Wait until the transmitter
reading is stable before starting the calibration.
3. Press MENU. The main menu screen appears. Choose Calibrate.
4. Choose Ozone.
5. Choose InProcess.
6. The screen at left appears for two seconds.
7. The screen at left appears. The number in the first line is the concen-
tration of ozone based on the previous calibration. Wait until the read-
ing is stable, then press ENTER.
8. The screen at left appears. Take a grab sample of the process liquid
and immediately press ENTER. The transmitter will store the present
sensor current and temperature and use those values in the calibration.
9. Immediately determine the ozone concentration in the sample.
10. Use the arrow keys to change the value in the second line to match the
results of the laboratory test. To return to the main display press MENU
then EXIT.
11. This screen appears momentarily if the sensitivity (sensor current divid-
ed by concentration) is much higher or lower than expected. The dis-
play then returns to the screen in step 5. See Section 15 for trou-
bleshooting.
Cal?
Ozone
Temp
Calibrate
Sim. PV
Program Display
Cal?
InProcess
Zero
Wait for
Stable reading.
Stable? 1.00ppm
Press enter.
Take sample;
Press enter.
Sample 1.00ppm
Cal
1
.00ppm
Calibration
Error
73