Instruction manual
49
MODEL XMT-A-FF/FI SECTION 7.0
PROGRAMMING THE TRANSMITTER
Measurement Temp
Security
>>
7.6.2 Procedure: Noise reduction
Calibrate Sim. PV
Program
Display
Ambient AC Power
60Hz
50Hz
Noise Rejection
ResetTransmitter >>
Noise Rejection
ResetTransmitter
>>
1. Press MENU. The menu screen appears. Choose Program.
2. Choose >>.
3. Choose Noise Reduction.
4. Select the frequency of the ambient AC power.
5. To return to the main display, press MENU then EXIT.
7.6 NOISE REDUCTION
7.6.1 Purpose
For maximum noise reduction, the frequency of the ambient AC power must be entered.
Measurement Temp
Security
>>
Calibrate Sim. PV
Program
Display
Load factory
settings?
Yes
No
1. Press MENU. The menu screen appears. Choose Program.
2. Choose >>.
3. Choose ResetTransmitter.
4. Choose Yes or No. Choosing Yes clears previous settings and calibrations and
returns the transmitter to the first quick start screen.
7.7 RESETTING FACTORY CALIBRATION AND FACTORY DEFAULT SETTINGS
7.7.1 Purpose
This section describes how to install factory calibration and default values. The process also clears all fault messages
and returns the display to the first quick start screen.
7.7.2 Procedure: Installing default settings