Instruction manual
MODEL XMT-A-FF/FI SECTION 7.0
PROGRAMMING THE TRANSMITTER
8a. If you chose total chlorine, select single or dual slope calibration. For the vast
majority of applications, dual slope calibration is unnecessary.
8b. Enter the time constant for the input filter.
8c. To return to the main display, press MENU then EXIT.
9a. If you chose monochloramine, enter the time constant for the input filter.
9b. To return to the main display, press MENU then EXIT.
Cl Cal Slope?
Single
Dual
7.4 MAKING TEMPERATURE SETTINGS
7.4.1 Purpose
This section describes how to do the following:
1. Choose temperature units (°C or °F).
2. Choose automatic or manual temperature correction for membrane permeability.
3. Choose automatic or manual temperature compensation for pH (pH settings apply to free chlorine only).
4. Enter a temperature for manual temperature compensation.
7.4.2 Definitions — oxygen, ozone, chlorine, and monochloramine
1. AUTOMATIC TEMPERATURE CORRECTION. Membrane-covered amperometric sensors produce a current directly
proportional to the rate the analyte (the substance being measured) diffuses through the membrane. The diffusion
rate is proportional to the concentration of analyte and the temperature. As temperature increases, membrane per-
meability increases. Thus, an increase in temperature will cause the sensor current to increase even though the
analyte level remained constant. A correction equation in the transmitter software automatically corrects for changes
in membrane permeability. In automatic temperature correction, the transmitter uses the temperature measured by
the sensor for the correction.
2. MANUAL TEMPERATURE CORRECTION. In manual temperature correction the transmitter uses the temperature
entered by the user for the membrane permeability correction. It does not use the actual process temperature. Do
NOT use manual temperature correction unless the measurement and calibration temperatures differ by no more
than about 2°C. Manual temperature correction is useful only if the sensor temperature element has failed and a
replacement sensor is not available.
7.4.3 Definitions — pH
1. AUTOMATIC TEMPERATURE COMPENSATION. The transmitter uses a temperature-dependent factor to convert
measured cell voltage to pH. In automatic temperature compensation, the transmitter measures the temperature and
automatically calculates the correct conversion factor. For maximum accuracy, use automatic temperature compen-
sation.
2. MANUAL TEMPERATURE COMPENSATION. In manual temperature compensation, the transmitter converts meas-
ured voltage to pH using the temperature entered by the user. It does not use the actual process temperature. Do
NOT use manual temperature compensation unless the process temperature varies no more than about ±2°C or the
pH is between 6 and 8. Manual temperature compensation is useful if the sensor temperature element has failed
and a replacement sensor is not available.
Input filter?
63% in
0
05sec
Input filter?
63% in
0
05sec
47