Instruction manual
MODEL XMT-A-FF/FI SECTION 7.0
PROGRAMMING THE TRANSMITTER
Manufacturer?
Rosemount
Other
Application?
Water/Waste
>>
units?
ppm
%sat ppb >>
units?
ppm
ppb
Pressure units?
mmHg
inHg atm >>
Use processPress
for %satn? Yes No
Process pressure
%sat: 7
60mmHg
Input filter?
63% in 0
05sec
Input filter?
63% in
0
05sec
Salinity parts/
thousand?
2
0.0
Measurement°
Temp
Security >>
Measurement type
Oxygen
Ozone >>
7.3.3 Procedure: Measurement
Calibrate Sim. PV
Program
Display
1. Press MENU. The menu screen appears. Choose Program.
2. Choose Measurement.
3. Choose Measurement type (oxygen, ozone, free chlorine, total chlorine, or
monochloramine).
4. The screen appearing next depends on the selection made in step 3.
a. If you chose oxygen, go to step 5a.
b. If you chose ozone, go to step 6a.
c. If you chose free chlorine, go to step 7a.
d. If you chose total chlorine, go to step 8a.
e. If you chose monochloramine, go to step 9a.
5a. Identify the manufacturer of the oxygen sensor: Rosemount or Other.
5b. Identify the application: water or wastewater, trace oxygen, or biopharm. Move
the cursor to >> and press ENTER to move to the next screen.
5c. Choose the units in which results are to be displayed: ppm, ppb, partialPress, or
%sat. Select >> to view the next screen. If you chose partialPress, the partial
pressure and the barometric pressure used in air calibration will be displayed in
the pressure units selected below.
5d. Choose pressure units: mm Hg, in Hg, atm, kPa, bar, or mbar.
5e. If percent saturation is to be calculated using the process pressure, choose Yes
and go to step 5f. If percent saturation is to be calculated using the barometric
pressure during air calibration, choose No. If you chose No, the screen changes
to the screen in step 5g.
5f. Enter the desired pressure.
5g. Enter the time constant for the input filter. See Sections 7.3.2 and 9.3.
5h. Enter the salinity in parts per thousand.
5i. To return to the main display press MENU then EXIT.
6a. If you chose ozone, select the units in which the ozone concentration is to be displayed.
6b. Enter the time constant for the input filter.
6c. To return to the main display, press MENU then EXIT.
45