Instruction manual
57
MODEL XMT-C SECTION 9.0
PROGRAMMING THE TRANSMITTER
9.4 CHOOSING AND CONFIGURING THE ANALYTICAL MEASUREMENT
9.4.1 Purpose
This section describes how to do the following:
1. Configure the transmitter to measure conductivity, resistivity, or TDS.
2. If resistivity was selected, one of the following temperature corrections must be selected:
a. Neutral salt
b. Slope
c. Cation conductivity
d. Raw/no temp correction
9.4.3 Procedure: Measurement.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Measurement.
3. Choose Conductivity, Resistivity, TDS, or Custom.
If you chose Conductivity or Resistivity, do steps 5 through 7.
If you chose TDS, no further measurement programming is required.
If you chose Custom, see Quick Start, step 8.
4. For Conductivity, select measurement units: (µS, mS)/cm or (µS, mS,
S)/m
5. Choose slope, cation, raw/none, or neutral salt for temperature cor-
rection.
6. If you chose Slope, enter the linear conductivity change in percent per
degrees (C or F). Press ENTER.
7. To return to the main display, press MENU followed by EXIT.
Calibrate Hold
Program
Display
Measure?
Cond
Resistivity >>
Unit? (
μS, mS
)
/cm
(μS, mS, S)/m
Temp Correction?
Slope Cation
>>
Temp Correction?
Raw/None
>>
Temp Correction?
NeutralSalt
>>
Measure?
TDS Custom >>
Temperature
Slope?
2
.00%/°C
Outputs Temp
Measurement
>>