Instruction manual

49
MODEL XMT-C SECTION 8.0
CALIBRATION — CONDUCTIVITY
8.1 ENTERING THE CELL CONSTANT
The cell constant should be entered:
• When the unit is installed for the first time
• When the probe is replaced
• During troubleshooting
All cell constants can be located on the cable label of the conductivity probe.
1. Press MENU. The main menu appears. Choose Calibrate. Press
ENTER.
2. Choose Conductivity. The screen at the left will appear. Press
ENTER.
3. Scroll right with the right arrow key. The screen at the left will appear.
4. Choose Enter Cell Constant. The screen at the left will appear.
5. Enter the actual cell constant as printed on the sensor tag attached to
the sensor cable.
NOTE
For sensors that show a "cal constant" on the label, the
actual cell constant can be calculated by adding 500 to
the cal constant, multiply this value by the nominal cell
constant, then divide the result by 1000.
6. Press ENTER. All conductivity readings will reference this sensor-spe-
cific cell constant.
7. Press EXIT once.
NOTE
The cell constant you have entered is changed after the
Standardizing Conductivity procedure is performed. For
inductive sensors and contacting sensors that only
show nominal cell constants, do not change it back to
the value as shown on the probe label.
Enter Cell Const
Temp Slope >>
Cal?
InProcess
Meter Zero >>
Cal?
Conductivity
Temperature
Cell Constant?
1
.0000/cm
If there is no cell constant on the label, calculate it
from the equation:
cell const = K
500 + cal const
1000
e
j