Instruction manual
29
FIGURE 4-12. ATEX Intrinsically Safe Installation for Model Xmt-C-FF
P
J. FLOCK
10/6/ 04
0mH
8.81 F
280mW
221mA
7.16V
300
300
800
240
240
13
Baseefa A
Li
Ci
DOES NOT DISSIPATE MORE THAN 1.3W.
10/6/ 04
10. THE ASSOCIATED APPARATUS MUST BE Baseefa APPROVED.
11. PROCESS RESISTIVITY MUST BE LESS THAN 10 OHMS.
9
WHEN INSTALLING THIS EQUIPMENT.
8. ASSOCIATED APPARATUS MANUFACTURER'S INSTALLATION DRAWING MUST BE FOLLOWED
GENERATE NOR STORE MORE THAN 1.5V, 100mA, 25mW OR A PASSIVE COMPONENT THAT
AS DEFINED IN ANSI/ISA
RP12.6 AND THE NEC, ANSI/NFPA 70. THEY CAN NOT
5. SENSORS WITHOUT PREAMPS SHALL MEET THE REQUIREMENTS OF SIMPLE APPARATUS
Li+ Lcable.
Ci+ Ccable;
Pmax OR Pi Po;
Imax OR Ii Isc, It OR Io;
Vmax OR Ui Voc,
Vt OR Uo;
FIELD DEVICE INPUT ASSOCIATED APPARATUS OUTPUT
WITH ASSOCIATED APPARATUS WHEN THE FOLLOWING IS TRUE:
7. THE ENTITY CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE APPARATUS
La, Lt OR Lo
Ca, Ct OR Co
MORE THAN 250 Vrms OR Vdc.
9. CONTROL EQUIPMENT CONNECTED TO ASSOCIATED APPARATUS MUST NOT USE OR GENERATE
LESS THAN 1.0 Ohm.
6. RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND AND EARTH GROUND MUST BE
3. I
NTRINSICALLY SAFE APPARATUS (MODEL XMT-P-FF, MODEL 375)
Pmax NOT GREATER THAN 0.9 W
Isc OR It NOT GREATER THAN 200 mA
Voc OR Vt NOT GREATER THAN 30 V
SUPPLY/SIGNAL TERMINALS TB2-1, 2 AND 3.
1. ANY SINGLE SHUN
T ZENER DIODE SAFETY BARRIER APPROVED BY CSA HAVING THE FOLLOWING OUTPUT PARAMETERS:
4. PREAMPLIFIER TYPE 23546-00, 23538-00 OR 23561-00 MAY BE UTILIZED INSTEAD OF THE MODEL XMT-P-FF
MAY BE ALSO USED. THE OUTPUT PARAMETERS SPEC
IFIED IN TABLE II ARE VALID FOR EITHER PREAMPLIFIER.
2. THE MODEL XMT-P-FF TRANSMITTER INCLUDES INTEGRAL PREAMPLIFIER CIRCUITRY. AN EXTERNAL PREAMPLIFIER
SPECIFIED IN TABLE I
THE CAPACITANCE AND INDUCTANCE OF THE LOAD CONNEC
TED TO THE SENSOR TERMINALS MUST NOT EXCEED THE VALUES
La Li (SENSOR) + Lcable.
WHERE Ca Ci (SENSOR) + Ccable;
23546-00 REMOTE PREAMPLIF IER.
TRANSMITTER INTEGRAL PREAMPLIFIER CIRCUITRY. A WEATHER RESISTANT ENCLOSURE MUST HOUSE THE TYPE
INDUCTANCE (La) WHICH CAN BE SAFELY CONNECT
ED TO THE APPARATUS. (REF. TABLES I, II AND III).
INCLUDING INTERCONNECTING WIRING, MUST BE EQUAL OR LESS THAN THE CAPACITANCE (Ca) AND
UNPROTECTED CAPACITANCE (Ci) AND INDUCTANCE (Li) OF THE INTRINSICALLY SAFE APPARATUS,
DELI
VERED BY THE ASSOCIATED APPARATUS (SAFETY BARRIER). IN ADDITION, THE MAXIMUM
EQUAL TO OR GREATER THAN THE VOLTAGE (Voc OR Vt) AND CURRENT (Isc OR It) WHICH CAN BE
THE VOLTAGE (Vmax) AND CURRENT (Imax)
OF THE INTRINSICALLY SAFE APPARATUS MUST BE
AND ASSOCIATED APPARATUS (SAFETY BARRIER) SHALL MEET THE FOLLOWING REQUIREMENTS:
C
B
A
EEx ia IIC T4
Baseefa04ATEX0214X
II 1 G
1180
(ZONE 0)
HAZARDOUS AREA
No modifications permitted
without the approval of
the Authorized Person
Related Drawing
Baseefa Certified Product
TB1-1 THRU 12
MODEL XMT-C-FF
XMTR
XMT-C-FF
MODEL
XMT-C-FF
XMT-C-FF ENTITY PARAMETERS
0.0
SUPPLY / SIGNAL TERMINALS TB2-1, 2 AND 3
321
23456 78 9 1011121
(SEE NOTES 1 & 9)
SAFETY BARRIER
GROUPS
GAS
OUTPUT PARAMETERS
(uH)
La
(uF)
Ca
IIB
IIA
TABLE I
IIC
Po
TABLE II
Uo
Io
PARAMETERS
OUTPUT
300
30
0.4
TABLE III
MODEL NO.
0
1.3
Vmax (Vdc)
Li (mH)
Ci (nF)
Pmax (W)
Imax (mA)
0.0
32
1.9
ENTITY PARAMETERS: REMOTE TRANSMITTER INTERFACE
Isc max OUT:uA
Voc max OUT: Vdc
Pamx IN: W
Imax IN:mA
Vmax IN: Vdc
375
MODEL NO.
1.0
20030
Li (mH)
Ci (uF)
FIELD COMMUNICATOR
ROSEMOUNT MODEL 375
TABLE III)
(SEE NOTE 2 AND
CLASS I AREA ONLY
INTERFACE FOR USE IN
REMOTE TRANSMITTER
SUITABILITY FOR DIVISION 2.
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR
DISCONNECT POWER BEFORE SERVICING.
TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES,
SENSOR
CONDUCTIVITY
WARNING-
24V TYP
30 VDC MAX FOR IS
POWER SUPPLY
UNSPECIFIED
WARNING-
UNCLASSIFIED AREA
LOAD
THIS DOCUMENT IS
CERTIFIED BY
REV
REV
REV
REV
REV
REV
REVISIONS NOT PERMITTED
W/O AGENCY APPROVAL
ATEX ZONE 0
MOD XMT-C-FF XMTR
B. JOHNSON
SCHEMATIC, INSTALLATION
Uniloc
9/ 30 /03
NOMINAL SURFACE FINISH 125
A
9065
6-30-05
1400276
1
1
SHEET OF
NONE
J. FLOCK
A
1400276
D
C
D
C
B
A
1
2
3
4
5
6
7
8
6
7
8
5
4
3
2
B
A
10-96
1
D
APPROVALS
.030
.XX
.XXX
.010
+
-
ANGLES
+
1/2
-
+
-
TOLERANCES
SOLID EDGE
CHK
REVISION
DATE
BYDESCRIPTION
ECO
LTR
NOTES: UNLESS OTHERWISE SPECIFIED
REVECO NO.
RELEASE DATE
PART NO.
ITEM
QTY
BILL OF MATERIAL
DESCRIPTION
DATE
THIS DWG CONVERTED TO
DRAWN
CHECKED
to those who may compete with Rosemount Analytical.
Rosemount Analytical, and is not to be made available
This document contains information proprietary to
TYPE
SCALE
TITLE
REV
DWG NO.
SIZE
ENGR APVD
PROJECT
MATERIAL
FINISH
MACHINED FILLET RADII .020 MAX
REMOVE BURRS & SHARP EDGES .020 MAX
DIMENSIONS ARE I N I NCHES
UNLESS OTHERWISE SPECIFIED
Rosemount Analytical,
Uniloc Division
2400 Barranca Pkwy