Instruction manual
MODEL 5081-C SECTION 7.0
PROGRAMMING
1. Press PRo G on the remote controller.
2. Press NEXT until the diSPLAy submenu appears. Press ENTER.
3. Press or
to display the desired measurement. Press ENTER to save.
4. tEMP will appear. Press or
to display the desired temperature reading, C or F.
Press ENTER.
5. OUtPUt will appear. Press or
to display the desired 4-20 output value, Cur or %.
Press ENTER.
6. COdE will appear. u sing the arrow keys on the remote control, enter the desired 3-digit
security code for accessing the Calibration, Program, and d iagnostic functions via the
Remote Control handheld. Press ENTER. The dISPLAy mnemonic will appear.
PRo GRAM
tYPE
EXiT ENTER
Conduc
PRo GRAM
dISPLAY
EXiT NEXT ENTER
7.4 DISPLAy
7.4.1 Purpose
This section describes how to do the following:
1. Configure the transmitter to measure conductivity, resistivity, or set up a custom curve
2. Set the temperature units to °C or °F
3. Set the output to current or percent of full scale
4. Enter a security code.
7.4.2 Definitions
1. MEASu REMENT. The transmitter can be configured to measure conductivity in mS/cm or resistivity in Mego hms, or
configured with a 3-5 point custom curve for special applications.
2. o u TPu T Cu RRENT. The transmitter generates a 4-20 mA output signal directly proportional to the conductivity or resis-
tivity of the sample. The output signal can be displayed as current (in mA) or as percent of full scale.
5. SECu RiTY Co d E. The security code unlocks the transmitter and allows access to all menus.
7.4.3 Procedure
COnduc Conductivity
rStvty Resistivity
CuSt Custom Curve
47