5.4 QUICK START FOR MODEL 5081-C-FF (FOUNDATION FIELDBUS)
1. o n the Remote, press PRo G, NEXT, ENTER.
2. u se the arrow buttons to select COnduc (conductivity), rStvty (resistivity), or CuSt (custom curve) mode. Press
ENTER. if you chose ConduC or rStvty , go to step 4.
3. if you selected CuST, you will see the Setup Custom screen. To move to the custom curve configuration menu, press
ENTER. You will automatically return to this same Setup Custom screen after configuration is complete. To continue
transmitter display programming, press NEXT in the Setup Custom screen.
4. u se the arrow keys to toggle temperature units between Celsius and Farenheit.
5. Press ENTER then RESET.
6. Press PRo G, ENTER.
7. u se the arrow key to toggle t AutO to o n or o FF to select using either the process temperature (tAutO = o n) or a
manual temperature (tAutO = o FF). Press ENTER. if you selected t AutO = o FF, you will be prompted to enter the
manual temperature; use the arrow keys, then press ENTER.
8. if you selected CondUC or rStvty in step 2, you will see a COMP (Temperature Compensation type) screen. u se the
arrow keys to select desired temperature compensation: LinEAr (linear), n SALt (neutral salt), CatIOn (cation con-
ductivity), or nOnE (raw or uncompensated conductivity). Press ENTER. if you are in LinEAR mode, you can now enter
a particular temperature slope (default is 2%/degC), then press ENTER to apply the slope.
9. Press RESET.
10. Press CAL, NEXT, NEXT, NEXT, NEXT, ENTER.
11. u se the arrow buttons to enter the cell constant of the sensor. Press ENTER.
12. if you are in rStvty , n SALt, Cation, or nOne (raw) mode, quick start is complete, proceed to step 20.
13. if you are in LinEAR mode, press EXiT, CAL, NEXT, ENTER.
14. Hold the sensor in air to zero. Press ENTER. PrOCESSIng will appear briefly. Press EXiT.
15. Place the sensor in a solution of known conductivity value. Press CAL, ENTER.
16. u se the arrow buttons to enter the current conductivity value of the solution. Press ENTER.
17. Press RESET.
To reset transmiter to factory default settings:
1. Press PRo GRAM, NEXT, NEXT, NEXT. The screen should say "dEFAULt". Press ENTER.
2. u se the arrow keys to toggle between nO (retain your configuration and calibration settings) and yES (restore factory
default settings to all variables).
3. Press ENTER, then EXiT.
MODEL 5081-C SECTION 5.0
DISPLAy AND OPERATION