Instruction manual
32
5.2 INFRARED REMOTE CONTROLLER (IRC) - KEy FUNCTIONS
The infrared remote controller is used to calibrate and program the transmitter and to read diagnostic messages.
See Figure 5-3 for a description of the function of the keys.
Hold the iRC within 6 feet of the transmitter, and not more than 15 degrees from horizontal to the display window.
FIGURE 5-3. Infrared Remote Controller.
RESET - Press to end the current oper-
ation and return to the  process display.
Changes  will  No T be  saved.  RESET
does not return the transmitter to factory
default settings.
CAL  - Press  to  access  the  calibrate
menu.*
PROG  - Press  to  access  the  program
menu.*
DIAG  - Press  to  view  diagnostic  mes-
sages.*
HOLD    - Press  to  access  the  prompt
that  turns  on  or  off  the  Hold  function.
Ho Ld  puts the transmitter in hold mode
and  sets  the  output  to  a  pre-pro-
grammed  value.  Press  RESET  to  exit
hold mode.
Editing Keys  - u se  the editing  keys to
change  the  value  of  a  flashing  display.
The left and  right  arrow keys move  the
cursor one digit at a time across a num-
ber.  The  up  and  down  arrow  keys
increase  or  decrease  the  value  of  the
selected  digit.  The  up  and  down  arrow
keys also  scroll the  display through the
items in a list.
* Pressing  CAL,  PROG, or  DIAG  causes
the  program  screen  to  appear  with  the
selected menu  (CALIBRATE, PROGRAM,
OR  DIAGNOSE) showing. See Figure  5-2.
The    first  sub-menu  (or  the  first  diagnostic
message)  also  appears.  Figure  5-4  shows
the complete menu tree.
ENTER - Press to advance from a sub-
menu to the first prompt under the sub-
menu. Pressing ENTER also stores the
selected  item  or  value  in  memory  and
advances to the next prompt.
NExT  - Press  to  advance  to  the  next
sub-menu.
ExIT - Press  to end the current opera-
tion.  The  transmitter  returns  to  the  first
prompt  in  the  present  sub-menu.
Changes will No T be saved.
MODEL 5081-C SECTION 5.0 
DISPLAy AND OPERATION