Instruction manual

77
MODEL XMT-P pH/ORP SECTION 7.0
PROGRAMMING THE TRANSMITTER
Outputs Temp
Measurement >>
Security HART
>>
7.8.2 Procedure: Noise reduction
Calibrate Hold
Program Display
Ambient AC Power
60Hz 50Hz
Noise Rejection
ResetTransmitter >>
Noise Rejection
ResetTransmitter >>
1. Press MENU. The menu screen appears. Choose Program.
2. Choose >>.
3. Choose >>.
4. Choose Noise Rejection.
5. Select the frequency of the ambient AC power.
6. To return to the main display, press MENU then EXIT.
7.7 MAKING HART RELATED SETTINGS
For more information refer to Section 6.0.
7.8 NOISE REDUCTION
7.8.1 Purpose
For maximum noise reduction, the frequency of the ambient AC power must be entered.
Outputs Temp
Measurement >>
Security HART
>>
Calibrate Hold
Program Display
Load factory
settings? Yes No
1. Press MENU. The menu screen appears. Choose Program.
2. Choose >>.
3. Choose >>.
4. Choose ResetTransmitter.
5. Choose Yes or No. Choosing Yes clears previous settings and calibrations
and returns the transmitter to the first quick start screen.
7.9 RESETTING FACTORY CALIBRATION AND FACTORY DEFAULT SETTINGS
7.9.1 Purpose
This section describes how to install factory calibration and default values. The process also clears all fault mes-
sages and returns the display to the first quick start screen.
7.9.2 Procedure: Installing default settings