Instruction manual

71
MODEL XMT-P pH/ORP SECTION 7.0
PROGRAMMING THE TRANSMITTER
7.3.3 Procedure: Configuring the Output
1. Press MENU. The menu screen appears. Choose Program.
2. Choose Output.
3. Choose Configure.
4. Choose Fault.
5. Choose Fixed or Live.
6. If you chose Fixed, the screen at left appears. Use the arrow keys to
change the fault current to the desired value. The limits are 4.00 to 22.00
mA. If you chose Live, there are no settings to make.
7. The screen at left appears. Choose mA/%.
8. Choose mA or percent. Percent means the display will show percent of
full scale reading.
9. The screen at left appears. Choose Damping.
10. Use the arrow keys to change the blinking display to the desired time con-
stant.
1. Press MENU. The menu screen appears. Choose Program.
2. Choose Output.
3. Choose Range.
4. Assign a value to the 4 mA output and press ENTER. Then assign a value
to the 20 mA output. Press ENTER. Use the arrow keys to change the
flashing display to the desired value.
Calibrate Hold
Program
Display
Display Output?
mA percent
Output Temp
Measurement°
>>
Configure? Fault
mA/% Damping
Configure? Fault
mA/% Damping
Configure? Fault
mA/% Damping
Set to value?
Fixed Live
Current Output
if Fault:22.00mA
Damping? 000−255
000 sec
Output range?
4mA 0.000ppm
Output? Test
Configure Range
Output Temp
Measurement°
>>
Output? Test
Configure Range
7.3.4 Procedure: Ranging the output
Calibrate Hold
Program Display