Instruction manual
43
MODEL XMT-C SECTION 9.0
PROGRAMMING THE TRANSMITTER
9.3.3 Procedure: Configuring the Output
1. Press MENU. The menu screen appears. Choose Program.
2. Choose Output.
3. Choose Configure.
4. Choose Fault.
5. Choose Fixed or Live.
6. If you chose Fixed, the screen at left appears. Use the arrow keys to
change the fault current to the desired value. The limits are 4.00 to 22.00
mA. If you chose Live, there are no settings to make.
7. The screen at left appears. Choose mA/%.
8. Choose mA or percent. Percent means the display will show percent of
full scale reading.
9. The screen at left appears. Choose Damping.
10. Use the arrow keys to change the blinking display to the desired time con-
stant.
1. From the main display, press MENU. The menu screen appears. Choose
Program.
2. Choose Output.
3. Choose Range.
4. Assign a value to the 4 mA output and press ENTER. Then assign a value
to the 20 mA output. Press ENTER. Use the arrow keys to change the
flashing display to the desired value.
Calibrate Hold
PPrrooggrraamm
Display
Display Ouput?
mmAA
percent
OOuuttppuutt
Temp
Measurement >>
Configure?
FFaauulltt
mA/% Damping
Configure? Fault
mmAA//%%
Damping
Configure? Fault
mA/%
DDaammppiinngg
Set to value?
FFiixxeedd
Live
Current Output
if Fault:
22
2.00mA
Damping? 000 255
00
00 sec
Output range?
4mA
00
.000μS/cm
Output? Test
CCoonnffiigguurree
Range
OOuuttppuutt
Temp
Measurement >>
Output? Test
Configure
RRaannggee
7.3.4 Procedure: Ranging the output
Calibrate Hold
PPrrooggrraamm
Display