MODEL XMT-C SECTION 7.0
CALIBRATION — TEMPERATURE
SECTION 7.0
CALIBRATION — TEMPERATURE
7.1 INTRODUCTION
The Calibrate Menu allows the user to calibrate the conductivity readings and temperature response of the
sensor.
7.2 CALIBRATING TEMPERATURE
7.2.1 Purpose
This procedure is used to ensure an accurate temperature measurement by the temperature sensor. It enables the
instrument to display process temperature accurately as well as to compensate for the effect of temperature on
the conductivity reading when the temperature in your process changes. The following steps should be performed
with the sensor in the process or in a grab sample near the operating temperature of the process.
7.2.2 Procedure
1. Check the instrument temperature reading (main display) to make sure the sensor has acclimated to the
process temperature. Compare the instrument temperature to a calibrated temperature reading device.
Proceed to the next step if the reading requires adjustment.
2. Press MENU. The menu screen appears. Choose Calibrate.
3. Choose Temperature.
4. To calibrate the temperature, change the number in the second line to
match the temperature measured in the process. Press ENTER.
5. Press MENU then EXIT to return to the main display.