Specifications
748414-A
August 1999
18
Heated Flame Ionization Detector Analyzer Module
Rosemount Analytical NGA 2000
3
STARTUP AND OPERATION
the network to locate and identify all components on the network. The user will have to
bind appropriate combinations of components after the startup sequence. (See section
3.3.)
7. Check the general health of the analyzer by reviewing the status of the Self Tests. All
“Pass” conditions should be obtained.
These test results can be found by selecting the following from the Main Menu:
Technical Level Configuration, Diagnostic Menus, Analyzer Module Diagnostics, Self
Test. All tested parameters should indicate "Pass."
Descriptions of the tests performed follow:
• EEPROM test - Checks the EEPROM on the Analysis Computer PCB.
• EPROM test - Checks the EPROM on the Analysis Computer PCB.
• RAM test - Checks the RAM on the Analysis Computer PCB.
• Power supply test - Verifies that all internal DC voltages are within the required
tolerances.
• Network test - Checks the internal network interface.
• 20 bit ADC test - Checks the 20-bit ADC on the Analysis Computer PCB by sending
a DC signal through the Preamp PCB and reading the signal back with the 20-bit
ADC.
• 12 bit ADC test - Checks the 12-bit ADC on the Analysis Computer PCB by sending
a DC signal and reading the signal back with the 12-bit ADC.
• Power Supply PCB test - Checks the presence of the Power Supply PCB by
activating the 3-way air solenoid.
• Safety PCB test - Checks the presence of the Safety PCB by sending a command
and reading it back.
• Case temperature test - Compares the temperature read between the Preamp
temperature sensor and the case temperature sensor. They must be within 10°C of
each other. This test sometimes fails if the case is opened. The sensor in the
Preamp will take longer to cool off since it is in an enclosure. Re-running the self-test
after thermal equilibrium will produce a positive result if the sensors are working
properly.
• Oven/Sample Temperature test - Compares the temperature read between the
sample temperature sensor and the oven temperature sensor. They must be within
50°C of each other.