Specifications

Heated Flame Ionization Detector Analyzer Module
Rosemount Analytical NGA 2000
748414-A
August 1999
7
INSTALLATION
2
F
IGURE
2-1. B
ACK
P
ANEL
C
ONNECTIONS
2.4.2 P
NEUMATIC
C
ONNECTIONS
WARNING: HIGH TEMPERATURES
The Sample In, Sample Bypass Out, and Burner Exhaust Out gases and fittings
can reach temperatures of up to 250
°
°°
°
C. Make connections to these fittings when
the oven heater is disabled or the module is powered down.
(See Figure 2-1) Connect inlet and outlet lines for sample, burner fuel and air, exhaust,
bypass, regulated air, and purge to appropriately labeled fittings on the rear panel. All
connections are 1/4-inch ferrule-type compression fittings except the PURGE AIR IN and
OUT connections, which are 3/8-inch compression fittings. The Burner Exhaust is a 3/8-inch
connection.
It is recommended that no connection be made to the PURGE AIR OUT port. If, however,
the analyzer's location requires interconnection with a venting system, the 3/8" O.D. line
should be kept as short as possible, and no longer than four feet.
CAUTION: POSSIBLE INSTRUMENT DAMAGE
No connection should be made to the PRESSURE RELIEF VALVE fitting. Doing
so may cause damage to the instrument.
Danger d'electrocution, d'explosion ou d'emanation de gaz
toxique. Se refere aux details inscrits surla face du module.
PURGE
AIR IN
PURGE
AIR OUT
PRESSURE
RELIEF
VALVE
BURNER
AIR IN
HEATED
SAMPLE IN
VENT TO SAFE AREA
SLOPE DOWNWARD 6
°
MINIMUM
FUEL
IN
BURNER
EXHAUST
OUT
REGULATE
D
MAXIMUM
INPUT
PRESSURE
HEATED
SAMPLE
BYPASS OUT
Possible electric shock, explosion or toxic gas hazard. See front of module
ATTENTION
FUEL: 50 psig(3445 hPa)
BURNER AIR: 50 psig (3445 hPa)
SAMPLE: 8 psig (551 hPa)
PURGE AIR: 20 psig (1378 hPa)
REG AIR: 25 psig (1722 hPa)
ATTENTION
WARNIN
G
WARNING
CHAUD !
!
!!
!
!
!!
!
HOT !