SPECIFICATIONS - PHYSICAL
CASE CLASSIFICATION:
General purpose for installation in weather-protected area
MAXIMUM SEPARATION
1600m (1 mile) from Analyzer Module to Platform
M
ATERIALS IN
C
ONTACT
W
ITH
S
AMPLE
Stainless steel and glass-filled Teflon
1
D
IMENSIONS
See Outline and Mounting Dimensions, Figure 2-5
W
EIGHT
15.9 kg (35 lbs.)
M
OUNTING
Horizontally, custom-installed in a panel
SPECIFICATIONS - GAS CONNECTIONS
S
AMPLE
I
N
:
1/4" O.D. tube fitting, stainless steel
R
EGULATED
A
IR
I
N
:
1/4" O.D. tube fitting, brass
B
URNER
A
IR
I
N
:
1/4" O.D. tube fitting, brass
F
UEL
I
N
:
1/4" O.D. tube fitting, stainless steel
P
URGE
A
IR
I
N
:
3/8" O.D. tube fitting, brass
P
URGE
A
IR
O
UT
:
3/8" O.D. tube fitting, brass
B
YPASS
O
UT
:
1/4" O.D. tube fitting, stainless steel
B
URNER
E
XHAUST
O
UT
:
3/8" O.D. tube connection, stainless steel (must slope downward
6
o
min. from horizontal)
Burner Exhaust, Bypass Out and Purge Air Out to be vented to atmospheric
pressure and to non-classified location in accordance with ANSI/NFPA-496
guidelines.
1