NGA 2000 MLT Hardware
TECHNICAL DESCRIPTION
1 - 42
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
Fig. 1-35: SIO-PCB with extension cards
1.10.8 DIO (Digital In-/Outputs)
Every platform can be equipped with max. 1 to 5 DIOs. Every MLT analyzer or every MLT analysis
module can be equipped with max. 1-2 DIOs (see table P-1, too)!
For maximum assembly with DIOs (5 or 2 DIOs resp.) there is no place for SIO PCB!
At mounting into a platform or an analyzer the DIO serves system functionallity of I/O´s and
supports all analysis modules of the NGA system.
Only the analysis module itself is supported at mounting into an analysis module.
To the programming we refer to the accompanying software instruction. The pin assigments are
described in chapter 21 of this instruction.
On DIO PCB are integrated:
8 digital Inputs,
5 - 30 VDC / 2,2 mA
Low level: 0,3 - 3,0 VDC / High level > 4 VDC
27 digital Outputs,
5 - 30 VDC / max. 500 mA
DIO (DIGITAL IN-/OUTPUTS)
Position of SIA module
Analog output 5 + 6
Position of SIA module
Analog output 3 + 4
Position of SIA module
Analog output 7 + 8
Position of
SIF module
RS232 or RS485