1.9.3 MLT 3 (gas purity measurement)
The necessary gas connections are marked analyzer specific (see section 5.3.and 5.4.3.1).
Different possibilities of internal layout and external tubing are shown in Fig. 1-30, depending on
instrument specific features.
Up to the internal pressure regulator the gas paths are designed as stainless steel tubings. Behind
the flow meter the gas paths are viton tubings.
All external fittings are swagelok
®
, stainless steel, 6/4 mm, 1/8 " or 1/4 ".
Fig. 1-30a: MLT 3 (gas purity measurement), gas path layout
(3 measuring channels with solenoid valve block option)
front side
rear side
1/8"
1/16"
1/16"
safety dust filter
threated plug 1/8"
6 mm
6 mm
SS capillary
SS tubing
SS
tubing
Viton tubing
Viton tubing
6 mm
6 mm
Viton tubing
tube piece
Viton tubing
analysis cell
analysis cell
PO2
Viton tubing
Viton tubing
span gas
zero gas
sample gas
pressure
sensor
gas outlet
1/8" 1/8"
INTERNAL GAS PATHS MLT 3 (GAS PURITY MEASUREMENT)