Instruction manual

NGA 2000 MLT Hardware
TECHNICAL DATA
20 - 11
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
01/2007
20.4 Voltage supply
20.4.1 Analyzers
20.4.1.1 MLT 1/4
Input 3-pole XLR- Flange (male), lockable
Voltage Supply 24 V dc (+/- 5 %) / 3 A (MLT 1) / 5 A (MLT 4)
[For ac operation {230/120 V} dc supply by options SL10, SL5 (both for rack
mounting only) UPS 01 T, or equivalent power supply]
20.4.1.2 MLT 2/3
Input terminal strips (MLT 2) / plug (MLT 3)
integrated power supply MLT 2: SL5 or SL10
MLT 3 UPS
Power Consumption MLT 2 max. 700 VA
Fuses MLT 2 (internal) T 6.3A/250 V (2 pcs.)
20.4.1.3 CAT 200
Connection via screw terminal blocks in
junction box, cross section: 4 mm
2
max.
Rated input voltage 115-230 V 50/60 Hz
Input voltage range 90 - 264 V , 47 - 63 Hz
Rated input power max. 380 W
20.4.1.4 Electrical Safety
Over-voltage category II
Pollution degree 2
Safety Class 2 for MLT 1/4 instruments.
1 for MLT 2/3 instruments
all I/O´s SELV voltage
optically isolated to electrical supply
VOLTAGE SUPPLY: ANALYZERS