Instruction manual

NGA 2000 MLT Hardware
TECHNICAL DATA
20 - 1
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
01/2007
20. Technical Data
Certifications EN 61326, EN 61010-1
(for measurment of not flammable gases or not
explosive gases resp. (< 50 % LEL).
NAMUR, CSA-C/US
*)
, C-Tick, PAC, BRML
Bigger concentrations requires supplementary GOSST: VNIIMS, Pattern (Belarussia)
protective measures !)
*)
MLT 2-NF (“Non-Flammable sample only”):
USA: Class I, Zone 2, Ex p II T4
Canada: AEx p II T4
PCB EXI 01 EC TYPE-EXAMINATION CERTIFICATION
TUEV Nord report No.: 98 ATEX 1341 X
FDA Test: 0-10 ppm CO (MLT1/3) TUEV Nord report No.: 98 CU 012
Suitability tests (MCERTS/TUEV/QAL 1): CO/NO/NO
2
/SO
2
/O
2
measurement
936/806017/A&D (MLT 1/ 2) TI Air, 13
th
BImSchV
936/806017/B&E (MLT 2/ 3 / 4) TI Air, 13
th
BImSchV, 17
th
BImSchV
- MLT 1 Sira MC 050051/00
- MLT 2 Sira MC 050052/01
- MLT 3/ MLT 4 Sira MC 050053/01
- MLT 1/ MLT 2 QAL 1 (EN 14181, EN ISO 14956)
- MLT 2/ MLT 3/ MLT 4 QAL 1 (EN 14181, EN ISO 14956)
20.1 Housing see dimensionsl skteches (Fig. 20-1 to 20-8)
Gas connections: Standard 6/4 mm PVDF
- Sample gas Option: 6/4 mm or 1/4", ss
- Reference gas MLT 1 max. 8 fittings
- Purge gas MLT 2 max. 6 fittings
MLT 3 max. 4 fittings
MLT 4 max. 6 fittings
CAT 200 max. 11 fittings
Weight (depending on configuration)
MLT 1 approx. 8 - 13 kg
MLT 2 (standard version) approx. 30 - 35 kg
MLT 3/4 approx. 13 - 18 kg
CAT 200 approx. 55 - 70 kg
HOUSING