Instruction manual
NGA 2000 MLT Hardware
REPLACEMENT AND CLEANING OF PHOTOMETRIC COMPONENTS
17 - 4
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
REMOVAL OF ANALYSIS CELLS
17.3 Cleaning of Analysis Cells and Windows
17.3.1 Removal of Analysis Cells
❍ Open the housing (cf. Section 15).
❍ Remove the photometer assembly out of analyzer housing (see Section 17.1).
a) For analysis cells of lengths 1 mm to 10 mm
❍ Remove the clamp (Fig. 17-3, Item 1).
❍ Remove the clamping collars and the filter cell with signal detector assembly.
Fig. 17-3: Photometer Assembly (1 mm to 10 mm analysisi cells)
b) For analysis cells of lengths 30 mm to 200 mm:
❍ Remove the clamp shown in Fig. 17-4 as Item 1.
❍ Remove the filter cell with signal detector assembly.
❍ Remove the clamp shown in Fig. 17-4 as Item 2.
❍ Remove the analysis cell body from the filter cell (chopper housing).
Fig. 17-4: Photometer Assembly (30 mm to 200 mm analysisi cells)
1
21
detector
adapter with
zero-point diaphragm