LIGHT SOURCE REPLACEMENT
Fig. 17-2: Chopper Housing with IR light sources
1 Light source hexagonal mounting screw
2 Light source with mounting flange
3 Temperature sensor (older versions)
4 Chopper Housing
4
17.2 Light Source Replacement (IR)
❍ Open the housing (cf. Section 15).
❍ Remove the photometer assembly out of analyzer housing (see Section 17.1).
❍ Remove the two light source hexagonal mounting screws (shown in Fig. 17-2 as Item 1).
❍ Remove the light source together with its mounting flange.
❍ Remove the mounting flange from the light source and position it on the new light source.
❍ Insert the new light source and flange in the same position as the old one.
❍ Insert and tighten the two light source hexagonal mounting screws (Fig. 17-2).
Then:
❍ Replace the photometer assembly (see Section 17.4).
❍ Perform the physical zeroing procedure (see Section 17.5).