Instruction manual

NGA 2000 MLT Hardware
OPENING OF HOUSING
15 - 15
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
Fig. 15-14: CAT 200,
Cast Iron housing, Interior View
1 Power supply unit
2 Filter pcb
3 Fuses for power supply unit
Repeat the steps in reverse order to assemble the housing.
When having assembled the dome perform a first test to ensure the front panel sensor keys can
be activated. If all sensor fields are accessible use the screws on top of the dome and extender
ring to fix both units (fig. 15-11).
If the sensor fields not accessible, see at section 15.4.3 to adjust the analyzer with magnetically
operated front panel.
CAT 200