Instruction manual
NGA 2000 MLT Hardware
OPENING OF HOUSING
15 - 13
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
15.4.2 Dome
To get access to internal components and the analyzer module the dome‘s components have to
be removed (fig. 15-11):
❍ Loosen the outer screw
❍ Turn the dome counterclockwise to remove it
❍ Loosen the inner screw
❍ Turn the extender ring counterclockwise to remove it
Fig. 15-11: CAT 200 Dome, fixing screws
❍ When dome is removed unscrew the knurled head screws at the analyzer module‘s sides
(fig. 15-12) and pull it to the front until the last screw locks into the detent (fig. 15-13).
Note!
Mark or remember the detent where the first knurled head screw was placed, to place it there
again when assembling the unit!
Otherwise it may happen the analyzer is not magnetically operable any more because the
distance between front panel and glass it to great.
CAT 200
Screws fixing dome
components