Instruction manual
NGA 2000 MLT Hardware
REMOVAL / REPLACEMENT OF COMPONENTS
12 - 6
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
FUSES
Fig. 12-3a: Fuses PCB LEM 01 (component side)
Fig. 12-3b: Fuses PCB LEM 02 (component side)
12.4.2 MLT 1 / 4
The fuses for 24 V dc Input are mounted on the PCB “LEM”.
❍ Remove PCB “LEM” (see Section 12.1).
❍ Take out and check the fuses (Fig. 12-3). Replace the fuse(s) if necessary.
Two pieces of spare fuses are on the PCB (Fig. 12-3).
❍ Insert PCB “LEM” (see Section 12.1) and reclose the instrument.
Network
Connectors
(RJ 45)
Fuse
Network
Connectors
(RJ 45)
Fuse
Spare Fuses
Spare Fuses