Instruction manual

NGA 2000 MLT Hardware
REMOVAL / REPLACEMENT OF COMPONENTS
12 - 4
Emerson Process Management GmbH & Co. OHG
Instruction Manual
90002929
07/2006
12.3.2 Replacement of Buffer Battery
Use the following procedure to replace the battery:
Remove the ACU 02 (see 12.3.1).
Remove the jumper “P23” (ACU02) or “P17” (ACU 01) for battery buffer (Fig. 12-1).
All data and compensation values entered by the user will be deleted (RAM-fail)!
Unsolder the battery from the soldering pins (see Fig. 12-1).
Solder the new battery (Ordering No. 03 765 180) to the soldering pins (see Fig. 12-1).
Verify polarity when soldering the new battery (Fig. 12-1) !
Do not short circuit battery !
After the replacement:
Re-insert the jumper for battery buffer (Fig. 12-1).
Install the ACU 02 (see 12.3.3).
12.3.3 Installation of ACU 02
Tilt the card ejector of the ACU 02 upwards and put in the card until the catch locks into
place.
Re-attach the cable (ACU 02 - AFP 01 connection) to the circuit board ACU 02.
Close the instrument (i.e., re-attach the front panel, see section 15.).
Switch on instrument.
Now all the data required by the user can be entered again, e.g. system parameters.
REPLACEMENT OF BUFFER BATTERY OF ACU 02